background image

3

3.04

The interior of the bath is constructed of

stainless steel and is designed for operation
with distilled water or water solutions, such as
water/ethylene glycol with a corrosion inhibitor
added. The body is made from galvanized steel
and is painted for added protection.  A stainless
steel gable cover is also provided with the bath.

3.05

The 230-volt units are identical in

appearance to the 120 volt units.

4.

PERFORMANCE DATA

4.01

The following table identifies the

specifications for the Coliform Incubator Bath.

Technical Specifications
Temperature Settings (factory preset)
  35 

o

C

  41.5 

o

C

  44.5 

o

C

  45.5 

o

C

Sensitivity (cover on)
  0.1

o

C

Uniformity (cover on)
± 0.2

o

C

Chamber Size
Length    14 inches (35 cm.)
Width

12 inches (30 cm.)

Height

8 inches (20 cm.)

Capacity

(1.5" from top) -
4.8 gallons (19.3L.)

5.

INSTALLATION

WARNING

INSTALLATION SHOULD BE COMPLETED
ONLY BY QUALIFIED PERSONNEL.

5.01

Location -

 The most uniform operating

conditions will be obtained by placing the bath
on a level surface in an area remote from drafts,
ventilating outlets, radiators, and other rapidly
changing ambient conditions.

Environmental Conditions-

This instrument is designed to operate safely
under the following conditions:

• Indoor Use Only
• Temperature: 5

° 

to 40

°

 C

• Maximum Relative Humidity: 80% for

temperatures to 22

°

C

• Maximum Altitude 2000 meters

Maximum performance is assured across the
following temperature range:

• 15ºC to 45ºC

5.02

Electrical Connections -

IMPORTANT

FOR PERSONAL SAFETY, THIS
APPARATUS MUST BE PROPERLY
GROUNDED.

1.

The power cord provided on this unit is

equipped with a three-prong (grounding) plug
which  mates with standard three-prong
grounding wall receptacle to minimize the
possibility of electric shock hazard from this
apparatus.  If in doubt the user should have the
wall receptacle and circuit checked by a qualified
electrician to make sure the receptacle can
provide adequate current and is properly
grounded.

2

.

Where a standard two-prong wall

receptacle is encountered, it is the personal
responsibility and obligation of the user to have
it replaced with a properly grounded three-prong
wall receptacle. 

Do not, under any

circumstances, cut or remove the third
(ground) prong from the power cord.  Do not
use a two-prong adapter plug.

Содержание 2860

Страница 1: ... Be Re quire d Pre ss or until d e sire d te mp e rature is d isplaye d Display will the n re turn to actual te mpe rature Pre ss ENTER Ne w Se t Point is acce pte d To Se le ct Te mp e rature Se le cte d Te mp e rature s Are Factory Prese t Pre ss ENTER calib ration is comp le te Pre ss or to match re ad out to the rmome te r Pre ss Calib rate b utton Place ce rtifie d trace ab le the rmome te r ...

Страница 2: ... DAMAGES DIRECT OR INCIDENTAL ARISING OUT OF OR RELATED TO THE USE OF THIS MANUAL For repair information or replacement parts assistance from the manufacturer call Technical Services using our tollfreetelephonenumber 800 438 4851 FAX 740 373 4189 NOTE THE 230V UNITS DESCRIBED IN THIS MANUAL WERE DESIGNED SPECIFICALLY FOR THE EUROPEAN MARKET AND ARE SUPPLIED WITH A EUROPEAN STYLE POWER CORD FOR DOM...

Страница 3: ... F NOV01 APR05 JUN06 SEPT08 APR09 Add caution acidic caustic substance newmanuals manufacturelocation consolidated with 3177691 230V removed CE symbol and declaration Revised water specifications Added UL required info 5 36100110 34001605 ECR23443 23496 ECR24808 ECR25448 BA 670 ...

Страница 4: ...UCTION 1 UNPACKING AND DAMAGE 1 GENERAL INFORMATION 2 PERFORMANCE 3 INSTALLATION 3 EXPLANATION OF CONTROLS 4 OPERATION 5 MAINTENANCE 6 TROUBLESHOOTING 7 REPLACEMENT PARTS LIST 10 ASSEMBLY SCHEMATIC DWGS 11 WARRANTY 15 ...

Страница 5: ......

Страница 6: ...packingmaterialuntilunitisput into service This merchandise was carefully packedandthoroughlyinspectedbeforeleaving our factory 2 02 Responsibility for safe delivery was assumedbythecarrieruponacceptanceofthe shipment therefore claimsforlossordamage sustained in transit must be made upon the carrier by the recipient as follows 1 Visible Loss or Damage Note any external evidence of loss or damage o...

Страница 7: ...RAL DEPOSITS AND TO PREVENT CORROSION USE ONLY DISTILLED WATER IN THE BATH WARNING USING CHLORINATED TAP WATER OR ADDITIVES THAT CONTAIN CHLORINE WILL VOID THE MANUFACTURER S WARRANTY SIMILARLY HIGH PURITY DEIONIZED WATER THAT DOES NOT FALL WITH IN RESISTIVITY RANGE OF 50K TO 1M OHM AND PH RANGE OF 7 TO 9 WILL ALSO VOID THE WARRANTY CONTACT TECHNICAL SERVICES WITH ANY QUESTIONS 3 03 The proportion...

Страница 8: ...fts ventilating outlets radiators and other rapidly changingambientconditions Environmental Conditions This instrument is designed to operate safely underthefollowingconditions IndoorUseOnly Temperature 5 to 40 C MaximumRelativeHumidity 80 for temperatures to 22 C Maximum Altitude 2000 meters Maximum performance is assured across the followingtemperaturerange 15ºC to 45ºC 5 02 Electrical Connectio...

Страница 9: ...l used to prevent thermalrunawayintheeventthattheprimary controlfails 6 03 Control Panel The Control Panel provides a digital readout of temperature readings heater ON indicator and pro vides keys for user to make their desired selections 1 LED Display Four digits are used to display the actual set and calibrated valuesfortemperature 2 Heater On Lamp The HEATER ON lamp is illuminated when power is...

Страница 10: ...will read actualbathtemperature 3 ToSelecttheOperatingTemperature Model 2860 has four user selectable temperature setpoints of 35 0 41 5 44 5 and45 5 C Toselectthedesiredoperating temperature presstheVordownWbutton until the desired temperature is displayed PressEnterbuttontoacceptthenewsetpoint Thedisplaywillflash indicatingthatthenew settinghasbeenentered Whentheflashing stops the display return...

Страница 11: ...ome discolored by iron rust use the following procedure to remove all traces of the rust and restore the stainless steel WARNING OBSERVE THE FOLLOWING SAFETY PRECAUTIONS USE HEAVY GLOVES OR OTHER ADEQUATE HAND PROTECTION WEAR GOGGLES OR OTHER ADEQUATE EYEPROTECTION ONLYWORKINAREAS WITH ADEQUATE VENTILATION Prepare a solution of 20 nitric and 1 5 hydrochloric acid if preferred a 2 to 5 solution of ...

Страница 12: ...it lamp is off If light is on turn High Limit control fully clockwise 3 Check temperature probe A Disconnect unit from electrical supply B Disconnect temperature probe connector J302 from CPU board C Apply a 5VDC to the temperature probe 5 to pin 1 of blue connector Ground to pin 3 of blue connector The connector has these numbers embedded on it D Place a voltmeter between pins 2 3 on temperature ...

Страница 13: ... See Paragraph 9 03 Step 6 5 Check TRIAC Q2 on power supply board A Turn power on B Let bath stabilize to desired setpoint value C Place an AC voltmeter between pins 1 5 on wire side of connector J105 on power supply board Voltmeter should fluctuate between 0 volts and anything less than 5 volts D If voltmeter does not show specified values in Step C replace power supply board 9 03 No Heat 4 Check...

Страница 14: ...pump 9 07 Replace CPU or Display Keyboard board 1 The software must be configured to match the Bath model To enter configuration mode press the upper left push button switch while applying power The upper left switch is labeled Calibrate This switch is used to enter configuration codes and steps through the different levels of configurations Immediately following power up with the switch depressed...

Страница 15: ...ssembly Display 3176679 Wire Harness 3177620 3173909 Temp Probe Assembly Plug w Jack 3167038 Probe Thermometer Holder 3177444 Bath O Ring Kit 3167129 Heater Kit 3167160 3175478 Heater Thermostat Bracket N A 3174465 High Limit Thermostat 3175458 Hi Limit Light Amber 3177575 3173907 On Off Switch 3175318 Line Cord 3178034 319481 Diffuser Plate 3163799 Gable Cover 3166219 Power Supply Fuse Mains N A ...

Страница 16: ...11 Stir Pump Assembly Stir Pump Extension On Off Power Switch Diffuser Plate 11 ASSEMBLY AND SCHEMATIC DRAWINGS FRONT SECTION VIEW ...

Страница 17: ...nsulation Certified Traceable Thermometer Holder Thermometer not supplied Temp Sensor Assembly Mounted Installed Here Stir Pump Knob Main Line Power Cord Power Line to Stir Pump Temp Probe Jack Plug probe end into jack Stir Pump Bracket ...

Страница 18: ...able Thermometer Holder Thermometer not supplied Temp Sensor Assembly Mounted Installed Here Stir Pump Knob Main Line Power Cord Receptacle Power Line to Stir Pump Temp Probe Jack Plug probe end into jack Stir Pump Bracket 230V UNIT BACK VIEW ...

Страница 19: ...14 TOP VIEW Temp Sensor Assembly Holder CertifiedTraceable ThermometerHolder Thermometernotsupplied Stir Pump Assembly Stir Pump Bracket Bath Pan Tub ...

Страница 20: ...15 ELECTRICAL DIAGRAM ...

Страница 21: ... warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Electron Corporation postage paid and replacement parts are shipped FOB destination THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLIED NO WARRANTIES OF MERCHANTABILITY...

Страница 22: ...m this warranty Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo postage paid and repla...

Страница 23: ......