Operating Range: 

Ambient Plus 5°C to 60°C

Temperature Uniformity:

+ 0.5°C

Temperature Stability: 

+ 0.5°C

Ambient Operating Range:  68-74°F (20-23°C)
Repeatability:

+ 0.5°C

Factory Calibrated to 37°C.

Heat-up time of unloaded dri-bath from a cold start to
37°C — 20 minutes.

Heat-up time for 40 each 10mm x 75mm or 12mm x
75mm tubes filled to heating block level with water placed
in a preheated block at 37°C — 10 minutes.

Designed for continuous service.

Intended Use

The Type 12200 Blood Bank Dri-Bath provides a con-
trolled dry heat environment for 10mm or 12mm test tubes
containing in vitro diagnostic specimens or samples for
qualitative or quantitative test procedures. Please refer to
the clinical laboratory method specified by the reagent
manufacturer or established by medical technology for
product application.

Function

The Blood Bank Dri-Bath provides incubating tempera-
tures in the range of Ambient Plus 5°C-60°C for 10mm or
12mm test tubes. The -60 units contain wells for 10mm
tubes.

Installation

Remove the incubator from the carton. Check the voltage
specifications and plug into the properly grounded recep-
tacle.

5

Introduction

Model

Volts

Amps

Watts

Hz

Wells

DB12215E, DB12215E-60

120

.4

47

50/60

40

Performance Characteristics & Specifications

Содержание 12200

Страница 1: ...Type 12200 Blood Bank Dri Bath Operation Manual and Parts List Model Voltage DB12215E 120 DB12215E 60 120 LT656X1 2 26 07 ...

Страница 2: ...use of this equipment Safety Information 3 Alert Signals 3 Warnings 3 Introduction 5 Performance Characteristics and Specifications 5 Intended Use 5 Function 5 Installation 5 Operation 6 10mm 60 Block Models 6 Calibration 6 12mm Block Models 7 Calibration 7 Maintenance 8 Cleaning 8 Parts List 9 Ordering Procedures 10 One Year Limited Warranty 12 2 Table of Contents ...

Страница 3: ... power supply prior to main tenance and servicing To avoid personal injury 1 Do not use in the presence of flammable or combustible materials Fire or explosion may result This device contains components which may ignite such materials 2 Caution Hot Surface Avoid Contact The block becomes hot while the unit is operating and will remain hot for some time after the unit is turned off 3 Refer servicin...

Страница 4: ...e located in the door the hearth collar in the chamber of the product or under the hot plate top Tests performed by the manufacturer indicate that there is no risk of exposure to dust or respirable fibers resulting from operation of this product under normal conditions However there may be a risk of exposure to respirable dust or fibers when repairing or maintaining the insulating materials or whe...

Страница 5: ...vironment for 10mm or 12mm test tubes containing in vitro diagnostic specimens or samples for qualitative or quantitative test procedures Please refer to the clinical laboratory method specified by the reagent manufacturer or established by medical technology for product application Function The Blood Bank Dri Bath provides incubating tempera tures in the range of Ambient Plus 5 C 60 C for 10mm or...

Страница 6: ... of the unit Turn the adjusting screw clockwise to increase temperature and counterclockwise to decrease temperature One com plete turn will change the temperature approximately 10 C Allow 10 minutes stabilization time between fine adjustments 6 Operation Note Do not place dri bath in a draft sun light or near equipment which emits heat or cold as this may result in fluc tuating temperatures Test ...

Страница 7: ...adjusting screw clockwise to increase temperature and counterclockwise to decrease temperature One com plete turn will change the temperature approximately 10 C Allow 10 minutes stabilization time between fine adjustments 7 OPERATION Note Do not place dri bath in a draft sun light or near equipment which emits heat or cold as this may result in fluc tuating temperatures Test tube liquid levels sho...

Страница 8: ...Cleaning Unplug incubator Use a moist cleaning cloth to clean the painted case and wells Dry thoroughly 8 Maintenance ...

Страница 9: ...art No Description No Req 120 Volt Models PC656X1A Control 1 EL656X1A Element 1 BK250X1A Block Heating 12MM 1 BK250X2A Block Heating 10 MM 60 Models 1 CR598X1 Cord Set 1 PLX97 Light Pilot 1 9 Parts List ...

Страница 10: ...c dealer from whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your request then contact our Customer Service Department at 563 556 2241 or 800 553 0039 Prior to returning any materials please contact our Customer Service Department for a Return Materials A...

Страница 11: ...11 ...

Страница 12: ...g use inconsistent with written operating instructions for the product mishandling contamination overheating modification or alteration of the product by any customer or third party or iii use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific prod ucts Heating elements because of their susceptibility to overheating and contamination must be re...

Отзывы: