Operating Range:
Ambient Plus 5°C to 60°C
Temperature Uniformity:
+ 0.5°C
Temperature Stability:
+ 0.5°C
Ambient Operating Range: 68-74°F (20-23°C)
Repeatability:
+ 0.5°C
Factory Calibrated to 37°C.
Heat-up time of unloaded dri-bath from a cold start to
37°C — 20 minutes.
Heat-up time for 40 each 10mm x 75mm or 12mm x
75mm tubes filled to heating block level with water placed
in a preheated block at 37°C — 10 minutes.
Designed for continuous service.
Intended Use
The Type 12200 Blood Bank Dri-Bath provides a con-
trolled dry heat environment for 10mm or 12mm test tubes
containing in vitro diagnostic specimens or samples for
qualitative or quantitative test procedures. Please refer to
the clinical laboratory method specified by the reagent
manufacturer or established by medical technology for
product application.
Function
The Blood Bank Dri-Bath provides incubating tempera-
tures in the range of Ambient Plus 5°C-60°C for 10mm or
12mm test tubes. The -60 units contain wells for 10mm
tubes.
Installation
Remove the incubator from the carton. Check the voltage
specifications and plug into the properly grounded recep-
tacle.
5
Introduction
Model
Volts
Amps
Watts
Hz
Wells
DB12215E, DB12215E-60
120
.4
47
50/60
40