3.2

OPTIONAL TRAPS

Optional traps (described in Section 2.1, Figure 2) must be maintained for optimal sys-
tem performance. The ANT100 utilizes neutralizing solutions which exhibit a color
change when depleted. Ammonia Neutralizing Solution (ANS121) changes from pink to
blue; Neutralizing Solution for Acids (NSA300TF) changes from blue to pale yellow
when depleted. Replace solution bottle when a color change is observed. If using solid
chemical cartridge with DTK120R, the DC120A cartridge for acids turns white to blue;
change when depleted. DC120R cartridge for radioactivity/organics has no color indica-
tor; replace monthly.

3.1

EMPTYING THE GLASS CONDENSATION FLASK (cont’d)

NOTE: 

If the Refrigerated Vapor Trap is not needed for several weeks, you 
may wish to shut it off between uses. Before switching on again, 
always remove the used glass vessel and replace with a clean, dry 
trap. Check the condition of the CryoCool; if a layer of water is 
visible under the CryoCool, remove with a pipette. Failure to follow 
these precautions may cause the glass vessel to break when the 
trap returns to operating temperature.

NOTE: 

Periodically check Flask Cover for signs of wear or perishing, and 
replace if necessary.

Содержание UVS400

Страница 1: ...A Thermo Electron business UVS400 UVS400A Universal Vacuum System INSTRUCTION MANUAL...

Страница 2: ...9 Section Section Title Warranty Page 6 0 Warranty and Liability 10 Disclaimer All statements information and data given herein are believed to be accurate and reliable but are presented without guara...

Страница 3: ...acuum pumps At the same time the UVS400 UVS400A is the most energy efficient and environmentally safe vacuum system available providing complete recovery of evaporated solvents Three major elements ar...

Страница 4: ...uum System on a stable level laboratory counter or heavy duty cart near a power outlet of the required voltage The bench depth should be at least 66 cm 26 inches The outlet must have a rating of at le...

Страница 5: ...he black rubber Flask Cover Press down flask and insulating cover so that the level of CryoCool rises around the flask and so that the flask its cover and insulation are seated to operating position T...

Страница 6: ...ast 30 minutes before using to ensure that it has reached its proper operating temperature of 50 C It is suggested that the trap be left on at all times When samples or gels are in place and the evapo...

Страница 7: ...lask Cover from flask leaving tubing attached to cover Withdraw the flask partially from the chamber and allow CryoCool to drain briefly Fully remove flask and insulating foam ring Avoid thermal shock...

Страница 8: ...ns white to blue change when depleted DC120R cartridge for radioactivity organics has no color indica tor replace monthly 3 1 EMPTYING THE GLASS CONDENSATION FLASK cont d NOTE If the Refrigerated Vapo...

Страница 9: ...0 Hz 6A 230 V 50 Hz 3 A 5 0 ACCESSORIES ANT100 Post trap Assembly ANS121 4 Ammonia neutralizing solution package or 4 bottles DTK120R Chemical Trap Kit for radioactive samples DC120A Disposable cartri...

Страница 10: ...epted without a Return Material Authorization RMA number issued by Thermo Savant Cost of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this man...

Страница 11: ...11...

Страница 12: ...45 3001 00 Rev A 2001 Thermo Savant 100 Colin Drive Holbrook NY 11741 4306 USA Tel 631 244 2929 1 800 634 8886 Fax 631 244 0606 www thermosavant com E mail savantec savec com A Thermo Electron busines...

Отзывы: