Thermo Savant UVS400 Скачать руководство пользователя страница 6

6

2.2 

CONNECTION TO EVAPORATION/DRYING EQUIPMENT

The UVS400 Universal Vacuum System is designed for use in single or multifunction
operation, with such instruments as the SpeedVac, Gel Dryer, Vacuum Oven, or Rotary
Evaporator. Use the tubing and fitting kit (TFK120) to arrange the components in a con-
venient, space saving configuration. If necessary, place the UVS400/UVS400A on a
shelf below the evaporation/drying components. Connect instrument(s) to the port on
the left side of the UVS400/UVS400A. Install manual bleeder valve (BV130; available
separately) in the tubing between the instrument and UVS400/UVS400A. This valve
can be used to apply vacuum to the instrument, and to bleed the system back to
atmosphere pressure at the end of a run.

3.0

OPERATION

A two-position power switch 

(OFF/ON)

is found on the right side panel. The system

should be switched 

ON

at least 30 minutes before using to ensure that it has reached

its proper operating temperature of -50 ˚C. It is suggested that the trap be left on at all
times.

When samples or gels are in place and the evaporation/drying component is activated,
switch the vacuum switch to 

ON

. This two-position switch 

(ON/OFF)

is located on the

lower right-hand side of the front panel.

VAPORNET
On the UVS400A, the two-position 

SOLVENT

switch (

VOLATILE/AQUEOUS

) is located

left-most on the front panel. If volatile, low boiling solvents (bp<95 ˚C) are being evapo-
rated, activate VaporNet by switching to 

VOLATILE

to maximize solvent recovery. If non-

volatile, high boiling solvents (bp>95 ˚C) are evaporated, select Aqueous (VaporNet off).

NOTE:

With 

VOLATILE

selected, occasional clicking will be heard from the 

system. This indicates that the VaporNet Controller is functioning properly.

When the concentration process is complete, switch the UVS400/UVS400A vacuum
pump 

OFF

. Shut off VaporNet when not in use by setting to 

AQUEOUS

.

Figure 2. a. ANT100 Post-trap Assembly b. DTK120R Chemical Trap Kit

a.

b.

Содержание UVS400

Страница 1: ...A Thermo Electron business UVS400 UVS400A Universal Vacuum System INSTRUCTION MANUAL...

Страница 2: ...9 Section Section Title Warranty Page 6 0 Warranty and Liability 10 Disclaimer All statements information and data given herein are believed to be accurate and reliable but are presented without guara...

Страница 3: ...acuum pumps At the same time the UVS400 UVS400A is the most energy efficient and environmentally safe vacuum system available providing complete recovery of evaporated solvents Three major elements ar...

Страница 4: ...uum System on a stable level laboratory counter or heavy duty cart near a power outlet of the required voltage The bench depth should be at least 66 cm 26 inches The outlet must have a rating of at le...

Страница 5: ...he black rubber Flask Cover Press down flask and insulating cover so that the level of CryoCool rises around the flask and so that the flask its cover and insulation are seated to operating position T...

Страница 6: ...ast 30 minutes before using to ensure that it has reached its proper operating temperature of 50 C It is suggested that the trap be left on at all times When samples or gels are in place and the evapo...

Страница 7: ...lask Cover from flask leaving tubing attached to cover Withdraw the flask partially from the chamber and allow CryoCool to drain briefly Fully remove flask and insulating foam ring Avoid thermal shock...

Страница 8: ...ns white to blue change when depleted DC120R cartridge for radioactivity organics has no color indica tor replace monthly 3 1 EMPTYING THE GLASS CONDENSATION FLASK cont d NOTE If the Refrigerated Vapo...

Страница 9: ...0 Hz 6A 230 V 50 Hz 3 A 5 0 ACCESSORIES ANT100 Post trap Assembly ANS121 4 Ammonia neutralizing solution package or 4 bottles DTK120R Chemical Trap Kit for radioactive samples DC120A Disposable cartri...

Страница 10: ...epted without a Return Material Authorization RMA number issued by Thermo Savant Cost of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this man...

Страница 11: ...11...

Страница 12: ...45 3001 00 Rev A 2001 Thermo Savant 100 Colin Drive Holbrook NY 11741 4306 USA Tel 631 244 2929 1 800 634 8886 Fax 631 244 0606 www thermosavant com E mail savantec savec com A Thermo Electron busines...

Отзывы: