Thermo Savant UVS400 Скачать руководство пользователя страница 7

7

3.1

EMPTYING THE GLASS CONDENSATION FLASK

During system operation, solvent vapors from the SpeedVac or other attached instru-
ment will collect in the glass condensation flask. The flask must be regularly emptied to
keep the entire system operating at peak efficiency. If the flask is not maintained, it may
become so full that sample drying rate is adversely affected. Another consequence of a
full vessel is that solvent vapors are more likely to contaminate the oil-sealed vacuum
pump bypass the trap and be released to the atmosphere. 

Empty the glass flask before it is half full. Often it is convenient and good practice to
change the flask at the end of the workday or workweek, no matter the depth of fill. For
aggressive solvents, you may wish to remove the flask at the end of each run for maxi-
mum protection of the system components. 

CAUTION:

The UVS400 and UVS400A reach low operating temperatures that can
cause severe damage to unprotected skin. Wear protective gloves 
and clothing when removing glass flasks. 

To remove the GCF400 for cleaning, bleed the system back to atmospheric pressure.
Remove rubber Flask Cover from flask, leaving tubing attached to cover. Withdraw the
flask partially from the chamber and allow CryoCool to drain briefly. Fully remove flask
and insulating foam ring. Avoid thermal shock by placing the flask on several thickness-
es of absorbent paper toweling and allow to come to room temperature. Insert a spare
GCF400, which is clean and dry, into the chamber. Cover with insulating foam ring, and
seal with rubber Flask Cover. Make sure rubber cover is seated firmly for a good vacu-
um seal. This easy system maintenance can be done in a matter of minutes; the power
(refrigeration) switch need not be shut off during this process. When the used flask has
defrosted, dispose of contents in an environmentally responsible manner. Clean and dry
flask for next use.

IMPORTANT:

Check the Glass Condensation Flask before each run and 
replace with clean flask if approaching half full. Immediately wipe
dry any CryoCool

®

Heat Transfer Fluid that falls on the rubber 

seal. Be sure Flask Cover is seated firmly on the flask.

NOTE:

Collected solvents may be purified for reuse or disposed of safely 
according to applicable regulations.

If working with radioactive samples, test condensate for presence of
radioactivity. If found to be radioactive, dispose of as radioactive
liquid according to all applicable regulations.

Содержание UVS400

Страница 1: ...A Thermo Electron business UVS400 UVS400A Universal Vacuum System INSTRUCTION MANUAL...

Страница 2: ...9 Section Section Title Warranty Page 6 0 Warranty and Liability 10 Disclaimer All statements information and data given herein are believed to be accurate and reliable but are presented without guara...

Страница 3: ...acuum pumps At the same time the UVS400 UVS400A is the most energy efficient and environmentally safe vacuum system available providing complete recovery of evaporated solvents Three major elements ar...

Страница 4: ...uum System on a stable level laboratory counter or heavy duty cart near a power outlet of the required voltage The bench depth should be at least 66 cm 26 inches The outlet must have a rating of at le...

Страница 5: ...he black rubber Flask Cover Press down flask and insulating cover so that the level of CryoCool rises around the flask and so that the flask its cover and insulation are seated to operating position T...

Страница 6: ...ast 30 minutes before using to ensure that it has reached its proper operating temperature of 50 C It is suggested that the trap be left on at all times When samples or gels are in place and the evapo...

Страница 7: ...lask Cover from flask leaving tubing attached to cover Withdraw the flask partially from the chamber and allow CryoCool to drain briefly Fully remove flask and insulating foam ring Avoid thermal shock...

Страница 8: ...ns white to blue change when depleted DC120R cartridge for radioactivity organics has no color indica tor replace monthly 3 1 EMPTYING THE GLASS CONDENSATION FLASK cont d NOTE If the Refrigerated Vapo...

Страница 9: ...0 Hz 6A 230 V 50 Hz 3 A 5 0 ACCESSORIES ANT100 Post trap Assembly ANS121 4 Ammonia neutralizing solution package or 4 bottles DTK120R Chemical Trap Kit for radioactive samples DC120A Disposable cartri...

Страница 10: ...epted without a Return Material Authorization RMA number issued by Thermo Savant Cost of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this man...

Страница 11: ...11...

Страница 12: ...45 3001 00 Rev A 2001 Thermo Savant 100 Colin Drive Holbrook NY 11741 4306 USA Tel 631 244 2929 1 800 634 8886 Fax 631 244 0606 www thermosavant com E mail savantec savec com A Thermo Electron busines...

Отзывы: