Used to pre-heat chamber prior to or between runs.
•
The chamber will always pre-heat to the set point on the display. (see page 3 on
how to modify setpoint temp.)
•
Cover must be left closed for pre-heat function to work. If left open, no heating of
chamber will occur.
(HINT) Pre-heating the SpeedVac
®
chamber will shorten the sample drying time by coun -
teracting evaporative cooling of the sample. Set temperature at 35 to 45˚C and leave
cover closed.
3.1
THE MANUAL RUN SPD101B
1.
Connect the unit to its required voltage.
2.
Turn the power switch to ON, located on the back of the unit. The coverlock
unlocks, allowing the top cover to be opened. (RED power light is ON.)
The display lights up, showing the default values:
TEMPERATURE:
45 ˚C (GREEN)
TIME:
0.00 HRS. (GREEN)
3.
Place sample tubes in rotor so load is balanced. Secure rotor with the supplied
knob (hand tight). Close cover.
4.
Using the up/down keys, set temperature between 45 ˚C - 80 ˚C.
5.
Press RUN/STOP button to initiate the run. (GREEN light is illuminated). The
cover locks and rotor starts spinning.
NOTE
:
If the cover is not closed the display will show “Lid” and the run will not
start.
6.
Wait approximately 1 minute for rotor to reach speed before applying vacuum to
chamber. This will stop samples from bumping!
7.
Open manual bleeder valve to apply vacuum to chamber.
8.
The timer counts up in 1 minute intervals. (GREEN decimal point blinks.)
9.
Temperature rises to set-point.
4
3 . 0
P R E - H E AT F U N C T I O N