background image

Used to pre-heat chamber prior to or between runs.  

The chamber will always pre-heat to the set point on the display. (see page 3 on
how  to modify setpoint temp.)

Cover must be left closed for pre-heat function to work. If left open, no heating of
chamber will occur.

(HINT) Pre-heating the SpeedVac

®

chamber will shorten the sample drying time by coun -

teracting evaporative cooling of the sample. Set temperature at 35 to 45˚C and leave
cover closed.

3.1

THE MANUAL RUN SPD101B

1.

Connect the unit to its required voltage. 

2.

Turn the power switch to ON, located on the back of the unit. The coverlock 
unlocks, allowing the top cover to be opened. (RED power light is ON.) 

The display  lights up, showing the default values: 
TEMPERATURE:

45 ˚C (GREEN)

TIME:

0.00 HRS. (GREEN)

3.

Place sample tubes in rotor so load is balanced. Secure rotor with the supplied 
knob (hand tight). Close cover.

4.

Using the up/down keys, set temperature between 45 ˚C - 80 ˚C. 

5.

Press RUN/STOP button to initiate the run. (GREEN light is illuminated). The 
cover locks and rotor starts spinning. 

NOTE

:

If the cover is not closed the display will show “Lid” and the run will not 

start.

6.

Wait approximately 1 minute for rotor to reach speed before applying vacuum to 
chamber. This will stop samples from bumping!

7.

Open manual bleeder valve to apply vacuum to chamber.

8.

The timer counts up in 1 minute intervals. (GREEN decimal point blinks.)

9.

Temperature rises to set-point. 

4

3 . 0

P R E - H E AT F U N C T I O N

Содержание SPEEDVAC SPD Series

Страница 1: ...SPEEDVAC CONCENTRATOR SPD SERIES BASIC SPD101B INSTRUCTION MANUAL A Thermo Electron business...

Страница 2: ...red Table of Contents Section Section Title Overview Page 1 0 Overview of the SPD101B Unit 1 1 1 Installation of the SPD101B Unit 2 1 2 How to Hook Up the SPD101B Unit 2 Section Section Title Control...

Страница 3: ...on 6 Right Side Panel 7 Chamber View Window 8 Vacuum Port 9 AC Socket 10 Main On Off Switch 11 Manual Cover Lock Release 12 Rear Access Panel 1 2 3 4 5 6 12 8 9 10 11 7 Figure 1 Front View Figure 2 Si...

Страница 4: ...9 Fax 631 244 0606 3 If the unit is part of a larger system use vacuum tubing supplied to hook up unit See figure 4 4 Vacuum clamps must be put on vacuum ports of pump and SPD unit See item 8 on figur...

Страница 5: ...view the preset parameters during the run 4 TEMPERATURE DISPLAY GREEN DISPLAY Indicates set temperature or actual temperature during pre heat or during a run from 45 C to 80 C or NO heat 5 RUN TIME D...

Страница 6: ...p cover to be opened RED power light is ON The display lights up showing the default values TEMPERATURE 45 C GREEN TIME 0 00 HRS GREEN 3 Place sample tubes in rotor so load is balanced Secure rotor wi...

Страница 7: ...ials 20 x 47mm v vials 1 dram vials 15 x 45mm 4ml 20 x 60mm v vials 18 x 52mm mini scintillation vials 28 x 60mm scintillation vials Microplates Deepwell Microplates 40 64 120 100 40 100 20 40 72 10 3...

Страница 8: ...choice please contact the Applications Support Specialists at Thermo Savant at 1 800 634 8886 or 631 244 2929 or Fax 631 244 0606 Part I Solvents suited for the SPD101B unit Ethanol Methanol Water Ace...

Страница 9: ...bsorbent towels Always wear gloves when cleaning and dispose of paper towel in appropriate designated refuse containers 3 Replace chamber seal if cracked Part number is 197 6020 00 4 Outside of unit c...

Страница 10: ...equential damages Warranty work is subject to our inspection of the unit No instruments equipment or accessories will be accepted without a Return Material Authorization RMA number issued by Thermo Sa...

Страница 11: ...YSTEM SpeedVac Concentrator SPD101B UVS400 Universal Vacuum Source ANT100 BV130 Bleeder Valve Vacuum Tubing COMPONENT HIGH VACUUM SYSTEM BV130 Bleeder Valve Flexible Vacuum Tubing DVG50 SpeedVac Conce...

Страница 12: ...SPD SERIES SPEEDVAC FAMILY SPD101B SPD111V SPD121P...

Страница 13: ......

Страница 14: ...97 3000 00 Rev B 2000 Thermo Savant 100 Colin Drive Holbrook NY 11741 4306 USA Tel 631 244 2929 1 800 634 8886 Fax 631 244 0606 www thermosavant com E mail savantec savec com A Thermo Electron busines...

Отзывы: