Thermo Savant SPEEDVAC SPD Series Скачать руководство пользователя страница 4

1 . 1

I N S TA L L ATION OF T H E SPD101B 

2

Savant

Vacuum Pump

Savant

Cold Trap

SPD101B

Manual 
Bleeder 
Valve

Vacuum Tubing

1.

Unpack unit from box and verify all parts match packing list.

2.

Read instruction manual carefully! If assistance is required, call Thermo Savant Customer 
Care at 1-800-634-8886, 631-244-2929, Fax:  631-244-0606.

3.

If the unit is part of a larger system, use vacuum tubing supplied to hook up unit. 
(See figure 4).

4.

Vacuum clamps must be put on vacuum ports of pump and SPD unit (See item 8 on figure 2,
and figure 4).

5.

If being hooked up to existing system, simply attach vacuum tubing to port on side of unit. 
(See item 8, figure 2-vacuum port).

CAUTION - Cold trap should be in line between vacuum source and SPD Unit.

1.2

How To Hook Up The SPD101B SpeedVac

®

System

Figure 4-Rear View of System

Clamp

Clamp

Содержание SPEEDVAC SPD Series

Страница 1: ...SPEEDVAC CONCENTRATOR SPD SERIES BASIC SPD101B INSTRUCTION MANUAL A Thermo Electron business...

Страница 2: ...red Table of Contents Section Section Title Overview Page 1 0 Overview of the SPD101B Unit 1 1 1 Installation of the SPD101B Unit 2 1 2 How to Hook Up the SPD101B Unit 2 Section Section Title Control...

Страница 3: ...on 6 Right Side Panel 7 Chamber View Window 8 Vacuum Port 9 AC Socket 10 Main On Off Switch 11 Manual Cover Lock Release 12 Rear Access Panel 1 2 3 4 5 6 12 8 9 10 11 7 Figure 1 Front View Figure 2 Si...

Страница 4: ...9 Fax 631 244 0606 3 If the unit is part of a larger system use vacuum tubing supplied to hook up unit See figure 4 4 Vacuum clamps must be put on vacuum ports of pump and SPD unit See item 8 on figur...

Страница 5: ...view the preset parameters during the run 4 TEMPERATURE DISPLAY GREEN DISPLAY Indicates set temperature or actual temperature during pre heat or during a run from 45 C to 80 C or NO heat 5 RUN TIME D...

Страница 6: ...p cover to be opened RED power light is ON The display lights up showing the default values TEMPERATURE 45 C GREEN TIME 0 00 HRS GREEN 3 Place sample tubes in rotor so load is balanced Secure rotor wi...

Страница 7: ...ials 20 x 47mm v vials 1 dram vials 15 x 45mm 4ml 20 x 60mm v vials 18 x 52mm mini scintillation vials 28 x 60mm scintillation vials Microplates Deepwell Microplates 40 64 120 100 40 100 20 40 72 10 3...

Страница 8: ...choice please contact the Applications Support Specialists at Thermo Savant at 1 800 634 8886 or 631 244 2929 or Fax 631 244 0606 Part I Solvents suited for the SPD101B unit Ethanol Methanol Water Ace...

Страница 9: ...bsorbent towels Always wear gloves when cleaning and dispose of paper towel in appropriate designated refuse containers 3 Replace chamber seal if cracked Part number is 197 6020 00 4 Outside of unit c...

Страница 10: ...equential damages Warranty work is subject to our inspection of the unit No instruments equipment or accessories will be accepted without a Return Material Authorization RMA number issued by Thermo Sa...

Страница 11: ...YSTEM SpeedVac Concentrator SPD101B UVS400 Universal Vacuum Source ANT100 BV130 Bleeder Valve Vacuum Tubing COMPONENT HIGH VACUUM SYSTEM BV130 Bleeder Valve Flexible Vacuum Tubing DVG50 SpeedVac Conce...

Страница 12: ...SPD SERIES SPEEDVAC FAMILY SPD101B SPD111V SPD121P...

Страница 13: ......

Страница 14: ...97 3000 00 Rev B 2000 Thermo Savant 100 Colin Drive Holbrook NY 11741 4306 USA Tel 631 244 2929 1 800 634 8886 Fax 631 244 0606 www thermosavant com E mail savantec savec com A Thermo Electron busines...