6
R70-32 (WEB)
REV(12-04) D
BATTERY LOCATION
AND REPLACEMENT
4", 6" AND 10" RECORDERS
To replace the battery, first open the recorder door.
The battery will be located in the upper right hand
corner of the unit. Refer to Figure 5 for the location of
the battery.
Note:
Use only NEW 9 volt alkaline
replacement batteries.
8" RECORDERS
CAUTION!
DO NOT TOUCH the
terminals of the transformer while the unit
is connected to the main AC power
supply. To avoid the risk of possible electric shock,
unplug or disconnect the recording unit from the main
power supply before changing the battery.
To replace the battery on an 8" recorder, first open the
recorder case door. Next, loosen the single screw
located on the right hand side of the chart plate and
open the hinged chart plate. The battery strap and
battery holder are located on the back of the chart
plate.
Note:
Use only NEW 9 volt alkaline
replacement batteries.
OPTIONAL ALARM/CONTROL RELAY
CONNECTION TO THE RELAY
CAUTION!
DO NOT TOUCH the
terminals of the transformer while the unit
is connected to the main AC power
supply. To avoid the risk of possible
electric shock, unplug or disconnect the recording unit
from the main power supply before attempting to
access the terminals of the relay. If the recorder is
installed with the battery back-up option, disconnect
the 9 Volt battery to avoid damaging the recorder and
draining the battery.
On 6" inch recording units mounted in a metal
enclosure, the chart plate on the recorder case is held
by four (4) screws. Unscrew and gently remove the
chart plate to expose the relay that is mounted to the
back of the chart plate. Otherwise, your recording unit
will have a hinged chart plate held in place by two (2)
screws located in the upper and lower right hand
corners of the chart plate. Remove these two screws
and open the hinged chart plate. On 8" recording units,
there is a single screw along the right hand edge that
secures the chart plate inside of the enclosure. Loosen
this screw and open the hinged chart plate to expose
the relay terminals. If you are not sure how to access
the relay terminals of your recording unit, please
contact C
O
B
EX
Recorders before continuing.
The relays that are used in this recording unit are
latching
type relays. That is, the contacts of the relay
will remain either closed or open (even when there is
no power applied to the recorder) until the relay is
pulsed with a signal from the recorder to change the
position of the contacts.
Refer to Figures 1B, 1C and 6B for the location of the
screw terminal connections for the relays.
Note:
A
RED
,
BLACK
and
WHITE
set of wires may already
be provided as an external connection to the relay's
terminal block. The terminal position NC will be
closed
when the pen is positioned to the right of the
control point and will be
open
when the pen is
positioned to the left of the control point.
The relays are rated for the following maximum
values:
2.0 AMP at 30V DC
0.6 AMP at 125V AC
0.6 AMP at 110V DC
Warning
: Damage to the recording unit may result if
the ratings for the relays are exceeded.
SETTING THE CONTROL POINT FOR
THE RELAY
To set the position of the pen on the chart that will
determine when the relay contacts will open or close,
follow these instructions:
Содержание 3940 Series
Страница 5: ...Thermo Electron Corporation iv Model 3940 Series Thermo Electron Corporation Preface ...
Страница 7: ......
Страница 19: ......
Страница 25: ......
Страница 29: ......
Страница 47: ......
Страница 51: ......
Страница 53: ......
Страница 54: ...Model 3940 Series 7 1 Thermo Electron Corporation Section 7 Refrigeration Schematics ...
Страница 55: ......
Страница 56: ...Model 3940 Series 8 1 Thermo Electron Corporation Section 8 Electrical Schematics ...
Страница 57: ...8 2 Model 3940 Series Thermo Electron Corporation Section 8 Electrical Schematics ...
Страница 58: ...Model 3940 Series 8 3 Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 59: ......
Страница 60: ...8 4 Model 3940 Series Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 61: ...Model 3940 Series 8 5 Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 62: ...8 6 Model 3940 Series Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 63: ......
Страница 76: ...11 R70 32 WEB REV 12 04 D Figure 7 6 Recorder Case Dimensions ...