Hold/Run Key
Pressing the Hold/Run key on the controller
silences the audible alarmand extinguishes the
humidity alarm indicator and the L4 indicator
on the controller. A4HI/A4LO flashes on the
lower display until the cabinet humidity
returns to the system setpoint.
Note
The humidity controller’s high and low
limits are factory-set at 100% and 0%.
Therefore, the system will go into the alarm state
when the humidity exceeds these percentages by
one percent. When operating the incubator near these high or low
humidity levels, frequent alarms may occur. This will require that the
controller’s high or low limit be reset to three or four percent over the high
limit or three or four percent under the low limit. Refer to the Watlow
User’s guide provided. Refer also to the factory configuration records
located at the end of Section 4 of this manual.
V
The Watlow temperature controller’s upper
numerical display shows the actual
temperature inside the incubator chamber.
The lower display shows the temperature
setpoint.
To raise or lower the setpoint, press the Up or
Down Arrow. Temperatures are set in 0.1°C
increments.
The Watlow humidity controller’s upper
numerical display shows the actual
humidity inside the incubator. The lower
display shows the humidity setpoint.
To raise the setpoint, press the Up or
Down Arrows. Humidity is set in one
percent increments.
2-4
Model 3940 Series
Thermo Electron Corporation
Section 2
Start-Up and Operation
Set the Operating
Temperature
Figure 2-5.
Hold/Run Key
Figure 2-6.
Changing Temp
Set the Operating
Humidity
Figure 2-7.
Changing Humidity
Содержание 3940 Series
Страница 5: ...Thermo Electron Corporation iv Model 3940 Series Thermo Electron Corporation Preface ...
Страница 7: ......
Страница 19: ......
Страница 25: ......
Страница 29: ......
Страница 47: ......
Страница 51: ......
Страница 53: ......
Страница 54: ...Model 3940 Series 7 1 Thermo Electron Corporation Section 7 Refrigeration Schematics ...
Страница 55: ......
Страница 56: ...Model 3940 Series 8 1 Thermo Electron Corporation Section 8 Electrical Schematics ...
Страница 57: ...8 2 Model 3940 Series Thermo Electron Corporation Section 8 Electrical Schematics ...
Страница 58: ...Model 3940 Series 8 3 Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 59: ......
Страница 60: ...8 4 Model 3940 Series Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 61: ...Model 3940 Series 8 5 Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 62: ...8 6 Model 3940 Series Thermo Electron Corporation Section 9 Electrical Schematics ...
Страница 63: ......
Страница 76: ...11 R70 32 WEB REV 12 04 D Figure 7 6 Recorder Case Dimensions ...