TITAN 1280 USER MANUAL
23
8-2 Real-time Temperature Data
Through serial output interface, the module output full frame temperature data
in real time. Refer to
Section
7-1-1 LVCMOS
and
Section 7-1-2 LVDS_H
for the
data format. The temperature is in Kelvin format.
There are two different kinds of data calculation format, for which the user can
consult the detailed usage or switch method. The two different data calculation
formats are as follows:
Format 1: Value =T x10
Format 2: Value =
{
T x10 - 7000, 0 ≤ Value ≤ 7300
T x15 - 3300, 7301 ≤ Value ≤ 16383