Testing a Cable
49
TV220 CableScout TDR
Test/Saved Waveform – Manual TDR
Display the current test waveform side by side with a saved waveform the same way as in
Auto TDR, described under “Test/Saved Waveform – Auto TDR” in this chapter. Figure 4-16
shows test and saved waveforms in Manual TDR.
Save/
Load
Vertical
Position
Vp
Setup
Main
Display
Test/Saved (Manual TDR)
Smoothing is 3 (Go to Setup – More settings to change)
Vp:
0.820
GAIN:
12 dB
PW:
5 ns
RL:
16 dB
:
17.9 ft
DIST:
35.2 ft
Marker On
Marker Off
feet
0
10
20
30
40
50
Figure 4-16: Test/Saved (Manual TDR)
The current test waveform is displayed in the top half of the waveform display and the saved
waveform is in the bottom half. The test waveform is scaled the same as the saved waveform
and displayed with a common cursor. Test settings are those of the saved waveform. Return-
loss values apply only to the current test waveform.
All test settings are from the saved waveform file, but you can change the Vp value on the
screen using
ST
. This does not affect the Vp value stored in the saved waveform file.
You can change the vertical position of the current test waveform.
Span
and
Pulse
softkeys are not available from this display.