background image

FL-B101D40PF Power factor (D0109) Manual

Page 1

This unit accepts an analog DC signal from a Power 

Factor transducer, ranging from 4-20mA, 0-1mA, 

2VDC, 5VDC, and 10VDC.

It provides a bargraph and digital display of the read-

ing in the format customary for Power Factor meters, 

with the bargraph centered at PF=1.0. 

Lead and lag are indicated by the bargraph extending 

upward (lead) or downward (lag) with cos(theta) scale 

markings from LAG 0.5 to 1.0 to LEAD 0.5. 

The digital display provides a corresponding numeric 

readout of the same data with LAG represented by 

negative sign.

• 

Choice of DC volts or DC current (header selectable ) 

•  Mono color (red or green), 101 segment high brightness bargraph. 

Vertical or optional horizontal format.

•  Red 4-digit LED display with a range of -0.50, 1, +0.50

 

Optional green digital display.

•  Auto-sensing AC/DC power supply. For voltages between 

 

85-265 V AC / 95-300 V DC (PS1) or 18-48 V AC / 10-72 V DC 

(PS2).

•  24 V DC excitation is available to power external 4/20mA

 

transmitters.

•  Provision to connect an external programming lockout switch. 
•  Provision for external DIM switch to reduce the brightest

 

display setting by 50%.

•  Optional NEMA-4 IP65, Clear dust & Splash proof front cover.
•  Automatic intelligent averaging, smooths noisy signals while 

 

providing a fast display response to real level changes.

•  UL Listed

General Features

Specifications

Index

FL-B101D40PF

Power Factor Bargraph Indicator

•  Bargraph center zero function.

•  Digital display blanking.

•  Four-level brightness control accessed by the   button 

and adjusted by the   button.

Software Features

Bargraph Center Point Display Mode Selection . . . 5

Case Dimensions . . . . . . . . . . . . . . . . . . . . . . . 9

Component Layout. . . . . . . . . . . . . . . . . . . . . . 7

Connector Pinouts . . . . . . . . . . . . . . . . . . . . . . 6

Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Controls and Indicators . . . . . . . . . . . . . . . . . . 2

Decimal Point and Brightness Selection . . . . . 5

Software Features  . . . . . . . . . . . . . . . . . . . . . .

1

Software Logic Tree . . . . . . . . . . . . . . . . . . . . . 3

Specifications. . . . . . . . . . . . . . . . . . . . . . . . . .

1

Two Point Analog Range Setting & Calibration . . . . 7

Two Point Digital Calibration Mode . . . . . . . . . 4

Digital Span Selection for Display . . . . . . . . . . 5

General Features . . . . . . . . . . . . . . . . . . . . . . .

1

Installation guidelines. . . . . . . . . . . . . . . . . . . . 9

Ordering Information  . . . . . . . . . . . . . . . . . . . 10

Panel Adaptor. . . . . . . . . . . . . . . . . . . . . . . . . .

8

Pin Descriptions . . . . . . . . . . . . . . . . . . . . . . . . 6

Programming Conventions. . . . . . . . . . . . . . . . 2

LEOPARD FAMILY

Input Specs: ��������������

• Header selectable current or voltage input 

signal

                                     • Current input: 4-20mA with shunt 100ohm; 

2mA range with shunt resistance 1kohm; 

(Voltage at full scale current input: 2V�)

                                     • Voltage input: 2V, 5V, 10V, input imped-

ance 1 Mohm

A/D Converter: ����������14 bit single slope
Accuracy: ������������������±(0�05% of r 2 counts)
Temp. Coeff�: �������������100 ppm/°C (Typical)
Warm up time: �����������2 minutes
Conversion Rate: ������10 conversions per second (Typical)
Digital Display: ����������4 digit 0.31" LED red (std), green (optn)
Bargraph Display: �����101 seg. bar red / 4 digit 4" LED red (std) 

green(optn), horizontal (optn)

Power Supply: �����������AC/DC Auto sensing wide range supply
  PS1 (std) ����������������85-265 VAC / 95-300 VDC, 50-400Hz 4.2W
  PS2 �������������������������18-48 VAC / 10-72 VDC, 50-400Hz 4�2W
Operating Temp�: ������0 to 50°C
Storage Temp: �����������–20°C to 70°C
Relative Humidity: ����95% (non condensing)
Case Dimensions: ����9/64 DIN (Bezel 36Wx144Hmm) 

Depth behind bezel (5�83") 148mm 
Plus (0�7”) 18mm for connectors 

Weight: �����������������������9�5 oz�, 12 oz when packed

Содержание Leopard Series

Страница 1: ...or Bargraph Indicator Bargraph center zero function Digital display blanking Four level brightness control accessed by the button and adjusted by the button Software Features Bargraph Center Point Display Mode Selection 5 Case Dimensions 9 Component Layout 7 Connector Pinouts 6 Connectors 6 Controls and Indicators 2 Decimal Point and Brightness Selection 5 Software Features 1 Software Logic Tree 3...

Страница 2: ...at an alternative logic branch will be followed when a particular option is present When two buttons are shown side by side and enclosed by a dotted line they must be pressed at the same time then released to go onto the next programming step If an X appears through a digit it means that any number displayed in that digit is not rel evant to the function being explained P Span 10000 P P When the a...

Страница 3: ...ut Lo cLo will occur This branch will only appear if the analog output option is installed DIGITAL SPAN SELECTION FOR ANALOG RANGE OUTPUT See Page 6 DIGITAL DISPLAY ON OFF See Page 5 BARGRAPH CENTER POINT DISPLAY MODE SELECTION See Page 6 TWO POINT ANALOG OUTPUT RANGE SETTING AND CALIBRATION See Page 7 Set the Decimal Point dp Select XX XX Bargraph Display Scale Lo Range Setting bLo Adjust to 50 B...

Страница 4: ...Digital Span Selection for Bargraph Display and the display flashes bhi and the previous setting ERROR Indicates Unsuccessful Calibration If the calibration was unsuccessful the display indicates Err the new calibration settings just entered will not take effect and the previously stored setting will remain The three most likely causes of an error during calibration are 1 The full scale and zero s...

Страница 5: ...ous brightness setting STEP D Set the Bargraph and Digital Display Brightness 1 Using the and buttons adjust the display to the desired brightness setting 4 is the brightest setting 2 Press the P button Display toggles between Cto and oFF Digital Span Selection For Bargraph Display Decimal Point and Brightness Selection Bargraph Center Point Display Mode Selection STEP G Bargraph Center Point Mode...

Страница 6: ...a unique plug and socket outline to prevent cross connection The main board uses standard right angled connectors Replacement 2 3 and 4 pin plug connectors are available see Accessories on page 10 Input Signal Pins 1 to 6 Pins 1 to 6 are reserved for the input signal conditioner See the data sheet for the selected input signal conditioner Pins 17 to 21 Rear Panel Switches Pin 19 Programming LOCK B...

Страница 7: ...ODULE 24V Exc IP07 Universal Process Input 2V 5V 10V 20V 200V 2mA 20mA Custom Custom 20mA 2mA 200V 20V 10V 5V 2V VOLTAGE CURRENT UNIVERSALPROCESS Processinput Requires Digital Calibration HI LO OFF ON 24VEXC 224C Input Range Header IP07 24V Exc on off Header Signal Type Header Voltage Current ...

Страница 8: ...l adapter enables modern DIN meters to fit in existing cutouts individually or stacked when replacing old 6 edgewise mechanical pointer meters Fits existing cut outs for 6 150 mm edgewise switchboard pointer meters from Crompton G E Westinghouse Yokogawa and most others Width 43 7 mm to 48 mm 1 72 to 1 89 Height 143 4 mm to 149 mm 5 62 to 5 88 Fits 6 Edgewise Pointer Meter Cut Outs Panel Adapter W...

Страница 9: ... IO board The adapter snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an existing 6 edgewise pointer meter cut out 100 0 25 50 75 100 0 25 50 75 SP1 SP2 SP3 SP4 Zero Span Zero Span Panel adaptor plates are available to retrofit most existing panel cutouts 31mm 1 22 These dimensions are increased by 2mm 0 08 when the metal surround case is installed For extra str...

Страница 10: ...odel number FL B101D40PF 144x36mm 101 Segment Bargraph 4 Digit Scale Factor Standard Options for this Model Number Order Code Suffix Description Display VRR Red LED Bargraph w 4 Digit Red DPM Vertical VGG Green LED Bargraph w 4 Digit Green DPM Vertical VGR Green LED Bargraph w 4 Digit Red DPM Vertical VRG Red LED Bargraph w 4 Digit Green DPM Vertical HRR Red LED Bargraph w 4 Digit Red DPM Horizont...

Отзывы: