background image

Texmate, Inc. Tel. (760) 598-9899  •  www.texmate.com

Page 6

DL-40JANUS-SG manual (d0116)

STEP A  Enter the Calibration Mode 

1)  Press the 

P

 and   buttons at the same time.

 

  Display toggles between [CAL] and [oFF].

 

2)  Press the   or   button. Display changes from [oFF] to [on].

 3) 

Press the 

P

 button. Display toggles between [CAL] and [out] input calibration.

Note: If at this point the display skips directly to toggle between [oFSt] (Manual) or [ZErO] 

(Auto) and the previous setting, the software is detecting that the optional analog output hard-

ware is NOT installed.
STEP B  Enter the Analog [oUT] Output Mode 

1)  Press the 

P

 button. Display toggles between [CLo] and an internal scale factor.

STEP C  Set or Calibrate the [CLo] Low Analog Output Range

 

1)  Select the voltage or current loop output header position on the output  

 

  module. (See Component Layout on page 9).

 

2)  Connect a multimeter to pins 16 and 17 on the output module. (See Rear    

  Panel Pinouts on page 8). Using the   and   buttons, adjust the analog    

  output to the desired low value as shown on the multimeter display. 

  cLo may be adjusted to any value from –0.3 mA to 17 mA (mA output  

 

  selected) or from –0.6 V to 8 V (volt output selected)

 3) 

Press the 

P

 button. Display toggles between [CHi] and an internal scale factor.

STEP D  Set or Calibrate the [CHi] High Analog Output Range

 

1)  Using the   and   buttons, adjust the analog output to the desired high    

  value as shown on the multimeter display. CHi may be adjusted to any value  

  from 17 mA to 21 mA (mA output  selected) or from 8 V to 10.3 V (volt out   

     put selected)

 

2)  Press the 

P

 button. The display exits the calibration mode and returns to    

  the operational display.

Note: Having established the Low and High range of the analog output, the digital span 

can now be selected which will set the two digital points between which the analog out-

put will occur. (See Digital Span Selection below).

STEP A  Enter the Decimal Point and Brightness Mode Through the Sub Menu 

[CAL] [oFF] 

1)  Press the 

P

 and   buttons at the same time. 

  Display toggles between [CAL] and [oFF]. 

2)  Press the 

P

 button. Display  shows previous [dP] selection.

STEP E  Set the Decimal Point 

1)  Using the   and  , adjust the display to the desired decimal point setting. 

2)  Press the 

P

 button. Display toggles between [br] and the previous [br] setting.

STEP F  Set the Display Brightness 

1)  Using the   and   buttons, adjust the display to the desired brightness  

  setting (4 is the brightest setting). 

2)  Press the 

P

 button. Display brightness changes to new setting and display  

  toggles between [AnHi] and the previous [AnHi] setting.

STEP G  Setting the Digital Span Point for Analog High Output 

1)  Using the   and   buttons, adjust the display to the desired digital value  

  which sets the point at which the selected analog high output range will occur. 

2)  Press the 

P

 button. Display toggles between [AnLo] and previous [AnLo] setting.

STEP H  Setting the Digital Span Point for Analog Low Output 

1)  Using the   and   buttons, adjust the display to the desired digital value  

  which sets the point at which the selected analog low output range will occur. 

2)  Press the 

P

 button. The display exits the calibration mode and returns to  

  the operational display.

Note: Any two digital scale points from –1999 to 9999 can be selected. The digital scale 

points for analog high and analog low can be reversed for reversed 20-4 mA output. The 

span of the digital scale can be as small as two counts however small spans cause the 16 

bit D to A to increment in stair case steps.

Two Point Analog Output Range Setting and Calibration

Decimal Point and Brightness Selection

Digital Span Selection for Analog Range Output

MAIN MENU

Operational Display

Sub-menu 

MODE

STEP A  

Calibration Mode

TWO POINT AUTO 

CALIBRATION or

MANUAL RESCALING MODE

See page 4 and 5

STEP B  

Calibration Mode

STEP C  

Calibrate Analog 

Output Lo

STEP D  

Calibrate Analog 

Output Hi

DECIMAL POINT AND 

BRIGHTNESS SELECTION

[X•XXX]

[XX•XX]

[XXX•X]

[XXXX•]

[XXXX]

[2]

[3]

[4]

STEP E  

Decimal 

Point (dp)

STEP F  

Display 

Brightness (br)

STEP G  

Analog 

High (Anhi)

STEP H  

Analog 

Low (AnLo)

Confirm

DIGITAL SCALE AND SPAN 

SELECTION FOR FULL SCALE 

ANALOG RANGE OUTPUT

Содержание DL-40JANUS-SG

Страница 1: ...utput Range Setting and Calibration 6 General Features Specifications Software Features Index LEOPARD FAMILY DL 40JANUS SG Two scaling methods MANUAL no external sig nal source required and AUTO requi...

Страница 2: ...by pressing and releasing the or buttons When there are more than two display selec tions they are shown in brackets below the first display and are also selectable by press ing and releasing the or b...

Страница 3: ...L POINT AND BRIGHTNESS SELECTION See Page 6 Setpoint 3 SP3 Setpoint 4 SP4 Valley Offset Scale Factor Calibrate Analog Output Lo Calibrate Analog Output Hi X XXX XX XX XXX X XXXX XXXX 2 3 4 Decimal Poi...

Страница 4: ...t STEP B Select Between Calibration of Input or Output Note If the analog output option is not present Step B is skipped and the program goes directly from Step A to Step C 1 Press the or button to se...

Страница 5: ...Point Analog Output Note If the analog output option is not present Step B is skipped and the program goes directly from Step A to Step C 1 Press the or button to select the display toggling from CAL...

Страница 6: ...s the P and buttons at the same time Display toggles between CAL and oFF 2 Press the P button Display shows previous dP selection STEP E Set the Decimal Point 1 Using the and adjust the display to the...

Страница 7: ...unts and does not include the deci mal point If you need the decimal value with decimal point you will need to read the register convert it to decimal read 40024 see below to determine decimal locatio...

Страница 8: ...ST2 Setpoint 2 Hysteresis setting 40019 0x0012 SINT16 R W 1999 9999 PRES3 Setpoint 3 setting 40020 0x0013 UINT16 R W 0 9999 HYST3 Setpoint 3 Hysteresis setting 40021 0x0014 SINT16 R W 1999 9999 PRES4...

Страница 9: ...P2 Delay on Make doM Delay Time Setting STEP H Set the SP2 Delay on Break dob Delay Time Setting STEP I Set the Hysteresis Setting for Setpoint 2 1 Using the and buttons adjust the display to the desi...

Страница 10: ...with external signal OFF 17 16 19 19 18 MANUAL rescaling ON AUTO scaling OFF ON SP1 NO SP1 NO SP2 NO SP2 NO Pins 14 and 15 AC DC Power Input Auto sensing AC DC power supply For voltages between 85 265...

Страница 11: ...air circulation 3 Separate input output leads from power lines to protect the meter from external noise Input output leads should be routed as far away as possible from contactors control relays tran...

Страница 12: ...71 1 8 DIN Cutout spacers 87 4mm 3 45 Various bezel colors are available Black is standard Prog SP3 SP4 SP2 SP1 For additional strength extra Mounting Slide Clips can be ordered and doubled up one beh...

Страница 13: ...se is made from fire retardant polycar bonate A metal surround case can be ordered to enhance the meter s fire retardant capabilities and also provide shielding against electromagnetic interference EM...

Страница 14: ...OA____ DL 40JANUS SG Add to the basic model number the order code suffix for each standard option required The last suffix is to indicate how many different special options and or accessories that you...

Отзывы: