When the rear Auto/Manual switch is in the Manual position, the meter can be rescaled without applying an external signal by 

changing the oFSt (offset) and SCAL (scale) parameters in the user menu.  A pre-calibrated Input Signal Conditioning module must 

be used.  The meter assumes that this calibration used a true zero input (0V from conditioner to meter) resulting in a display value 

of 0 and a maximum input signal span (2V from conditioner to meter) resulting in a display value of the SCAL value set in the user 

menu.  The linear equation is y = mx + b where m = (SCAL-0)/(2-0) and b = oFSt.

The default value of SCAL is 2000, but it may be set to any value between -1999 and +9999.

The internal Signal Span is limited to 3 V DC between- 1 V DC to+ 2 V DC. Any outputs from an Input Signal Conditioning module 

that exceed these limits will cause the meter to indicate  overrange regardless  of the Digital Display  Span scaled.  For example, 

an ID01 DC Volts Input Signal Conditioner can read +/-2VDC, but in the DL-40 will show under-range for a -1.2V input.

STEP A  Enter the Calibration Mode 

1)  Press the 

P

 and   buttons at the same time.

 

  Display toggles between [CAL] and [oFF].

 

2)  Press and hold the   or   button.

 

  Display changes from [oFF] to [on].

 

3)  Press the 

P

 button. Display toggles between [CAL] and [out].

STEP B  Select Between Calibration of Input or Output

Note: If the analog output option is not present, Step B is skipped and the 

program goes directly from Step A to Step C.

 

1)  Press the   or   button to select the display toggling from [CAL]  

  to [inPt].

 

2)  Press the 

P

 button. Display toggles between [oFSt] and the  

  previous offset setting.

STEP C  Set the Offset on the Digital Display

 

1)  Using the   and   buttons, adjust the digital display to the  

  desired offset. This is the reading that the meter will display for a  

  zero input.

 

2)  Press the 

P

 button. Display toggles between [SCAL] and the  

  previous Scale factor.

STEP D  Set the Scale factor on the Digital Display

 

1)  Using the   and   buttons, adjust the meter display to the  

  desired Scale factor. The default value is 2000, for which a 2V  

  input will read 2000. If the scale factor is changed the display will  

  change proportionately. Therefore if the Scale factor is changed  

    to 1000 then for the same 2V input the display would read 1000.

 

 

2)  Press the 

P

 button. 

The Digital Calibration Procedure Mode is Now Complete

The menu branches to the DECIMAL POINT AND BRIGHTNESS SE

LECTION, (see page6) and the display flashes [dP] and the previous 

decimal point selection.

Manual Rescaling Mode

Manual Rescaling Procedure

MAIN MENU

Operational Display

Sub-menu 

MODE

STEP A  

Calibration

Mode

STEP B  

Calibration

Mode

DECIMAL POINT AND 

BRIGHTNESS SELECTION

See Page 6

STEP C  

Offset

STEP D  

Scale

Factor

DECIMAL POINT AND 

BRIGHTNESS SELECTION

See Page 6

Decimal Point 

(dp)

TWO POINT ANALOG 

OUTPUT RANGE SETTING 

AND CALIBRATION

SEE PAGE 6

Содержание DL-40JANUS-PROCESS

Страница 1: ...l Features Specifications Software Features Index LEOPARD FAMILY DL 40JANUS PROCESS Two scaling methods MANUAL no external sig nal source required and AUTO requied an exter nal signal source Optional...

Страница 2: ...ed by pressing and releasing the or buttons When there are more than two display selec tions they are shown in brackets below the first display and are also selectable by press ing and releasing the o...

Страница 3: ...MAL POINT AND BRIGHTNESS SELECTION See Page 6 Setpoint 3 SP3 Setpoint 4 SP4 Valley Offset Scale Factor Calibrate Analog Output Lo Calibrate Analog Output Hi X XXX XX XX XXX X XXXX XXXX 2 3 4 Decimal P...

Страница 4: ...out STEP B Select Between Calibration of Input or Output Note If the analog output option is not present Step B is skipped and the program goes directly from Step A to Step C 1 Press the or button to...

Страница 5: ...wo Point Analog Output Note If the analog output option is not present Step B is skipped and the program goes directly from Step A to Step C 1 Press the or button to select the display toggling from C...

Страница 6: ...ress the P and buttons at the same time Display toggles between CAL and oFF 2 Press the P button Display shows previous dP selection STEP E Set the Decimal Point 1 Using the and adjust the display to...

Страница 7: ...counts and does not include the deci mal point If you need the decimal value with decimal point you will need to read the register convert it to decimal read 40024 see below to determine decimal locat...

Страница 8: ...HYST2 Setpoint 2 Hysteresis setting 40019 0x0012 SINT16 R W 1999 9999 PRES3 Setpoint 3 setting 40020 0x0013 UINT16 R W 0 9999 HYST3 Setpoint 3 Hysteresis setting 40021 0x0014 SINT16 R W 1999 9999 PRE...

Страница 9: ...e SP2 Delay on Make doM Delay Time Setting STEP H Set the SP2 Delay on Break dob Delay Time Setting STEP I Set the Hysteresis Setting for Setpoint 2 1 Using the and buttons adjust the display to the d...

Страница 10: ...g with external signal OFF 17 16 19 19 18 MANUAL rescaling ON AUTO scaling OFF ON SP1 NO SP1 NO SP2 NO SP2 NO Pins 14 and 15 AC DC Power Input Auto sensing AC DC power supply For voltages between 85 2...

Страница 11: ...es with 4 Independent 400V 210mA DC only SSRs DL 40JANUS Series 29 28 27 25 26 24 23 22 Options Order Code OR32 4A 4A 4A 4A 4A 4A OR34 4A 4A 4A 4A 4A 4A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules w...

Страница 12: ...XC FullyUserScalable PIN 2 PIN 1 PIN 3 _ Exc On Off Header SPAN Pot HI LOW SPAN RANGE Header ZERO Pot ZERO ADJUST Header SPAN ADJUST Header ZERO OFFSET RANGE Header Decrease Increase Decrease Increase...

Страница 13: ...ar 15 Turn Potentiometer 100 Counts 100 Counts 0 ZERO Potentiometer Pot If provided the ZERO pot is always to the left of the SPAN pot as viewed from the rear of the meter Typically it enables the inp...

Страница 14: ...4 7mm 0 19 DIN Cutout Spacer 2mm 0 08 Connector Socket Prog SP1 SP1 SP2 SP4 SP3 SP5 SP6 Prog SP1 SP1 SP2 SP4 SP3 SP5 SP6 Prog SP1 SP1 SP2 SP4 SP3 SP5 SP6 When extra panel mounting tightness is requir...

Страница 15: ...case is made from fire retardant polycar bonate A metal surround case can be ordered to enhance the meter s fire retardant capabilities and also provide shielding against electromagnetic interference...

Страница 16: ...RELAY OUTPUT OPTIONS ACCESSORIES OA____ DL 40JANUS PROCESS Add to the basic model number the order code suffix for each standard option required The last suffix is to indicate how many different spec...

Отзывы: