background image

Page 6

CL-B101D40-TCRTD Manual (d0099)

Example of Using the Center Point Bargraph Display Mode

with a Unipolar Input

If the meter's full scale range is set to 5000 counts, the midpoint would 

be 2500 counts. If a signal of 2500 counts is applied only one segment at 

the 2500 count mark will light up. If a signal of 4000 counts is applied the 

segments between the center segment (2500 counts) and the 4000 count 

mark light up.

If a signal of 1000 counts is applied, the segments between the center seg-

ment (2500 counts) and the 1000 count mark will light up.

Example of Using the Center Point Bargraph Display Mode

with Bipolar Signal Inputs

The 

meter

 may also be calibrated to display symmetrical bipolar signals such 

as ± 1 V or ± 10 V. When the center point display mode is selected, it will 

then function as a center zero meter.  When positive signals are applied, the 

bar will go up from the center point, and when negative signals are applied, 

the bar will go down from the center point.

STEP E  Selecting the [Anhi] Digital Value for Analog High Output

 

1)  Using the   and    buttons, adjust the display to the desired 

digital value at which the [chi] Calibrated Analog High output will 

occur.  For digital readings outside the digital span selected, the 

analog output will linearly rise above the value set for chi, up to the 

maximum analog output capability.  However, the analog output will 

not go lower than the calibrated value set for cLo (see below).

 

2)  Press the 

P

 button. Display toggles between [AnLo] and 

previous [AnLo] setting.

STEP F  Selecting the [AnLo] Digital Value for Analog Low Output

 

1)  Using the   and   buttons, adjust the display to the desired  

digital value at which the [cLo] Calibrated Analog Low output will 

occur.  For Digital readings outside the Digital Span selected, the 

analog output will not go lower than the calibrated value set for 

cLo.

 

2)  Press the 

P

 button. The display toggles between [cto] and 

[oFF].

Note: Any two digital span points from –1999 to 9999 can be selected. The digital 

values for [Anhi] analog high and [AnLo] analog low can be reversed to provide a 20 to 

4mA output. The digital span selected can be as small as two counts, when using the 

analog output to function as a Control or Alarm Driver.  Small digital spans will cause 

the high resolution 16 bit D to A to increment digitally in stair case steps.

Digital Span Selection for Analog Range Output

Digital Span Selection for Analog Range Output

Bargraph Center Point Display Mode Selection

STEP G  Bargraph Center Point Mode Selection (See example above)

 

1)  To select bargraph center point mode, press the   or    

  button. Display changes from [oFF] to [on].

 

2)  Press the 

P

 button. Display toggles between [diSP] and  

 

     [on] or [oFF].

STEP H  Digital Display ON/OFF Selection

 

1)  To set the display to [oFF], press the    or   button. 

 

     Display toggles between [diSP] and [oFF].

 

2)  Press the 

P

 button. The display exits the calibration mode  

  and returns to the operational display. Only the bargraph  

  display is on and the digital display is off.

If the digital display is selected to be off, pressing any button to 

make programming changes or to view setpoints activates the digital display. 

When the procedure is complete, the digital display will then automatically 

switch off.
The Display/Bargraph settings are now complete. 

Operational Display

STEP G

STEP H

See Example of Bargraph 

Center Point Display Mode 

Selection Above

From Digital Span Selection

for Analog Range Output Above

or Direct From Decimal Point

and Brightness Selection Page 5

if Analog Output Option is Not Present

BARGRAPH CENTER POINT
DISPLAY MODE SELECTION

STEP E

STEP F

This branch will only

appear if the analog 

output option is installed

To Bargraph Center

Point Display Mode

Selection Below

From Decimal Point and

Brightness Selection

See Page 5

DIGITAL SPAN SELECTION

FOR ANALOG RANGE OUTPUT

-1V

+1V

0V

-1V

+1V

0.800

-0.800

0V

-1V

+1V

0V

Center Point

Single Bar Lit

Center Point

As signal

increases

Digital display

Digital display

Center Point

As signal

decreases

Digital display

0

5000

5000

2500

2500

4000

0

5000

2500

1000

0

Center Point

Single Bar Lit

Center Point

As signal

increases

Digital display

Digital display

Center Point

As signal

decreases

Digital display

Unipolar Signal Input

Bipolar Signal Input

Содержание CL-B101D40-TCRTD

Страница 1: ...or below Lo each setpoint Digital display blanking Decimal point setting Four level brightness control LEOPARD FAMILY Input Specs Depends on Input signal conditioner A D Converter 14 bit single slope...

Страница 2: ...direction indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are sho...

Страница 3: ...999 seconds Pump PUM Hysteresis hYSt Hysteresis hYSt Hysteresis hYSt Setpoint 3 SP3 Setpoint 4 SP4 Operational Display MAIN MENU BRIGHTNESS AND SETPOINT VIEW MODE Bargraph Center Point Display Selecti...

Страница 4: ...E STEP A Calibration Mode STEP B Calibration Mode To Digital Span Selection for Bargraph Display Page 5 cAL out iP branch will only appear if the analog output option is installed otherwise cAL on goe...

Страница 5: ...between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph Display Bargraph does not light up for Input Signals up to...

Страница 6: ...igh and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Smal...

Страница 7: ...nd 18 see Rear Panel Pinouts on page 10 Using the and buttons adjust the analog output to the desired low value as measured on the multimeter cLo may be adjusted to any value from 0 3 mA to 18 mA mA o...

Страница 8: ...Press the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value...

Страница 9: ...Bargraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color se...

Страница 10: ...p Form A relays An analog output module is also shown as installed The CL B101D40 TCRTD uses plug in type screw terminal con nectors for all input and output connections The power supply connections p...

Страница 11: ...5A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog Output Module Mai...

Страница 12: ...the back panel back into the case Apply power to the meter 4 Enter the program mode and select the type of thermocouple J K R T the resolution 0 1 or 1 and the display units C or F See Page 4 of the d...

Страница 13: ...10 72VDC Input Modules Partial List See www texmate com IT10 Thermocouple J K R T Selectable C F 1 0 1 IT11 RTD 100 Pt Selectable 3 4 wire C F 1 0 1 385 392 IT15 RTD 1000 Pt Selectable 3 4 wire C F 1...

Отзывы: