CL-B101D40-TCRTD Manual (d0099)

STEP A  Enter the Calibration Mode

 1) 

Press the 

P

 and   buttons at the same time. 

Display toggles between [CAL] and [oFF].

 

2)  Press the   or   button. Display changes from [oFF] 

to [on].

 

3)  Press the 

P

 button. 

  

STEP B

  If the Display toggles between [CAL] and [out] the 

optional Analog  Output hardware is installed. In which 

case select [CAL] [out]. Display toggles between [thEr] 

or [rtd] depending on whether a IT10 Thermocouple 

Input Module or a IT11 RTD Input Module are detect-

ed by the soft ware. If no optional output hardware is 

installed the menu will skip directly to STEP C.

STEP C  Sensor Type Selection

 

1) Using the   and   buttons, adjust the display 

to the desired sensor type.

 

2) Press the 

P

 button. Display toggles between 

[rES] resolution select and previous [rES] setting.

STEP D  Set the Resolution

 

1) Using the   and   buttons, adjust the display 

to the desired resolution [rES] value.

 

2)  Press the 

P

 button. Display toggles between 

[dEG] and previous [dEG] selection.

STEP E  Selection of Degree C or Degree F

 

1) Using the   and   buttons, adjust the display to 

either ˚C or ˚F.

 

2) Press the 

P

 button. 

Calibration Procedure

MAIN MENU

Operational Display

Sub-menu 
MODE

STEP A  Calibration

Mode

STEP B  Calibration

Mode

To Digital Span Selection for

Bargraph Display Page 5

cAL out/iP branch will only appear if 

the analog output option is installed, 

otherwise cAL on goes directly to step c.

To Digital Span Selection

for Bargraph Display

See Page 6

To Two Point Analog

Output Range Setting

and Calibration

See Page 7

[K]

[r]

[t]

If RTD module is installed

this menu appears

3 or 4 wire 

RTD input 

selection

RTD select type 
P385 or P392

If Thermocouple module is installed

this menu appears

Resolution select

Thermocouple select

Select J, K, R or T

STEP C

STEP D

STEP E

1

°

0.1

°

Degree select

°

C

°

F

THERMOCOUPLE OR RTD TYPE,

RESOLUTION AND DEGREE C OR F SELECT

Input Sensor

Reso-

lution

°

C Range

°

F Range

J T/C

K T/C

R T/C

T T/C

100

Ω

 RTD (385 curve)

100

Ω

 RTD (392 curve)

-120  to  760

°

C

-120.0  to  530.0

°

C

-120  to  1370

°

C

-120.0  to  530.0

°

C

0  to  1760

°

C

0    to  530.0

°

C

-120  to  400

°

C

-120.0  to  400.0

°

C

-200  to  800

°

C

-199.9    to  530.0

°

C

-200  to  800

°

C

-199.9    to  530.0

°

C

-200  to  1400

°

F

-199.9  to  999.9

°

F

-200  to  2500

°

F

-199.9  to  999.9

°

F

32  to  3210

°

F

32.0  to  999.9

°

F

-200   to  750

°

F

-199.9  to  750.0

°

F

-200   to  1470

°

F

-199.9  to  999.9

°

F

-200   to  1470

°

F

-199.9  to  999.9

°

F

Sensor Range Table

1

°

0.1

°

1

°

0.1

°

1

°

0.1

°

1

°

0.1

°

1

°

0.1

°

1

°

0.1

°

Thermocouple type or RTD type selection mode

Input Module Calibration Procedure

See page 12 for the Calibration instructions of each Input Module type.

Содержание CL-B101D40-TCRTD

Страница 1: ...or below Lo each setpoint Digital display blanking Decimal point setting Four level brightness control LEOPARD FAMILY Input Specs Depends on Input signal conditioner A D Converter 14 bit single slope...

Страница 2: ...direction indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are sho...

Страница 3: ...999 seconds Pump PUM Hysteresis hYSt Hysteresis hYSt Hysteresis hYSt Setpoint 3 SP3 Setpoint 4 SP4 Operational Display MAIN MENU BRIGHTNESS AND SETPOINT VIEW MODE Bargraph Center Point Display Selecti...

Страница 4: ...E STEP A Calibration Mode STEP B Calibration Mode To Digital Span Selection for Bargraph Display Page 5 cAL out iP branch will only appear if the analog output option is installed otherwise cAL on goe...

Страница 5: ...between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph Display Bargraph does not light up for Input Signals up to...

Страница 6: ...igh and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Smal...

Страница 7: ...nd 18 see Rear Panel Pinouts on page 10 Using the and buttons adjust the analog output to the desired low value as measured on the multimeter cLo may be adjusted to any value from 0 3 mA to 18 mA mA o...

Страница 8: ...Press the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value...

Страница 9: ...Bargraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color se...

Страница 10: ...p Form A relays An analog output module is also shown as installed The CL B101D40 TCRTD uses plug in type screw terminal con nectors for all input and output connections The power supply connections p...

Страница 11: ...5A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog Output Module Mai...

Страница 12: ...the back panel back into the case Apply power to the meter 4 Enter the program mode and select the type of thermocouple J K R T the resolution 0 1 or 1 and the display units C or F See Page 4 of the d...

Страница 13: ...10 72VDC Input Modules Partial List See www texmate com IT10 Thermocouple J K R T Selectable C F 1 0 1 IT11 RTD 100 Pt Selectable 3 4 wire C F 1 0 1 385 392 IT15 RTD 1000 Pt Selectable 3 4 wire C F 1...

Отзывы: