50
9-4. AC Voltage
Note
The specifications are only applicable for sinusoidal signals with amplitudes
greater than 5% of the Full Scale reading, excluding the DL-2052 which must
have amplitudes greater than 10.0mV when using a range of 100.000mV.
(*) Input > 450V only for 30sec, < 200V for 20 ~ 45Hz
Rate
Range
Resolution
Full Scale
Input Impedance
S
100.000mV
1µV
120.000mV
1.1M
Ω in parallel
with approx.100pF
1.00000V
10µV
1.20000V
10.0000V
100µV
12.0000V
100.000V
1mV
120.000V
750.00V(*)
10mV
750.00V
M
100.00mV
10µV
120.00mV
1.0000V
100µV
1.2000V
10.000V
1mV
12.000V
100.00V
10mV
120.00V
750.0V(*)
100mV
750.0V
F
100.0mV
100µV
120.0mV
1.000V
1mV
1.200V
10.00V
10mV
12.00V
100.0V
100mV
120.0V
750V(*)
1V
750V
Rate
Range
Accuracy (rdigits)
20~45Hz
45~10kHz
10k~30kHz
S
100.000mV
1% + 100
0.2% + 100
1.5% + 300
1.00000V
1% + 100
0.2% + 100
1% + 100
10.0000V
1% + 100
0.2% + 100
1% + 100
100.000V
1% + 100
0.2% + 100
1% + 100
750.00V(*)
1% + 100
0.2% + 100
1% + 100
M
100.00mV
—
0.2% + 40
1.5% + 80
1.0000V
—
0.2% + 40
1% + 40
10.000V
—
0.2% + 40
1% + 40
100.00V
—
0.2% + 40
1% + 40
750.0V(*)
—
0.2% + 40
1% + 40
F
100.0mV
—
0.2% + 5
1.5% + 10
1.000V
—
0.2% + 5
1% + 5
10.00V
—
0.2% + 5
1% + 5
100.0V
—
0.2% + 5
1% + 5
750V(*)
—
0.2% + 5
1% + 5
9-5. DC Current
Note
mA range protected with a 2A fuse
10A range protected with a 12A, 600V fuse
10A only for 30 seconds
Rate
Range
Resolution
Full Scale
Accuracy
(r digits)
S
10.0000mA
0.1µA
12.0000mA
0.05%+15
100.000mA
1µA
120.000mA
0.05%+5
1.0000A
100µA
1.2000A
0.2%+5
10.0000A
100µA
10.0000A
0.2%+5
M
10.000mA
1µA
12.000mA
0.1%+6
100.00mA
10µA
120.00mA
0.1%+3
1.000A
1mA
1.200A
0.2%+3
10.000A
1mA
10.000A
0.2%+3
F
10.00mA
10µA
12.00mA
0.1%+2
100.0mA
100µA
120.0mA
0.1%+2
1.00A
10mA
1.20A
0.2%+2
10.00A
10mA
10.00A
0.2%+2