User Manual for the FLUXMEDICARE® V2 System
-
IFU-FLUXMEDV2 - V.1.0 - 07/2020
Page 48 / 84
4.4.4.
Starting treatment
•
After initialisation, the warm-up of the light source laser is carried out.
Treatment can be started from 15°C. This step may take several minutes.
•
If the optical connector for the TEXTLIGHTX-V2 luminous bands is not
connected, the screen tells the user to connect it to the intermediate unit
and the red error warning light comes on. Once the connection has been
made, press the START/PAUSE button to continue.
•
If the count key is not connected, the screen tells the user to connect it to
the intermediate unit with a sound signal and the red error warning light
on. Once the connection has been made, press the START/PAUSE button
to continue.
•
Once the initialisation has finished, the screen displays the number of
treatments remaining on the count key.
•
To start the treatment, press the “START/PAUSE” button.
•
The treatment starts. The blue indicator light comes on to show that the
treatment is in progress.
The remaining treatment time is shown on the screen (in h = hours: min =
minutes: s = seconds).
•
The luminous band light up.
4.4.5.
Pausing
•
When the laser is emitting, the blue indicator light is on, the screen displays the remaining treatment
time.