Software Overview
39
SBOU123A – March 2012 – Revised September 2016
Copyright © 2012–2016, Texas Instruments Incorporated
XTR108EVM-USB Evaluation Board and Software Tutorial
5.3.7
Verify Calibration
After successfully completing the manual calibration procedure, the minimum-scale, mid-scale, and
maximum-scale
Percent Error
indicators should be less than or equal to 0.1%. See
for an
example of a successful post-calibration screen.
Figure 26. Manual Calibration Tab, Post-Calibration
5.3.8
Reset Calibration
The
Reset Calibration
button can be clicked at any time in order to cancel the current calibration and
reset the
Manual Calibration
tab to the initial state. At this point, a new calibration can be performed with
either the current model file or a newly loaded model file.