TEST AFE5807/08
•
If each channel has the output as shown in the following illustration, proceed with the next
step; otherwise, contact the TI FAE to troubleshoot the problem.
•
On the AFE5808 GUI, change Test Pattern from Ramp to None for the next step.
8.2
Step 2: Single Tone FFT
•
Select the Single Tone FFT page at the TSW1250 GUI.
•
Connect Channel 1 of the AFE5807EVM to a signal generator through an LP filter. If an LP filter is not
present, the result will not be good.
•
Set the amplitude of the signal generator to -25dBm.
•
Set the frequency of the signal generator to 2.00439453M to match the GUI.
•
Change the window option to Hanning. This is because the input signal and the onboard CMOS clock
are noncoherent.
•
Press the Capture button to get the test result.
•
Repeat for Channel 2 through Channel 8.