Turn On the TSW14J57EVM Power and Connect to the PC
www.ti.com
10
SLAU701 – May 2017
Submit Documentation Feedback
Copyright © 2017, Texas Instruments Incorporated
Setup Procedure
NOTE:
1. The LMKCLK frequency must be equal to the frequency of the DEVCLK.
2. Ensure that the DEVCLK and LMKCLK sources are frequency-locked using a common
10-MHz reference to ensure functionality. Frequency locking the input signal generator to
the other generators can also be done if coherent sampling is desired.
3. Do not turn on the RF output of any signal generator at this time.
4. When using the ADC in single-input mode, the device uses both edges of DEVCLK for
sampling.
3.6
Turn On the TSW14J57EVM Power and Connect to the PC
1. Turn on the power switch of the TSW14J57EVM.
2. Connect a mini-USB cable from the PC to the TSW14J57EVM.
3. If this is the first time connecting the TSW14J57EVM to the PC, follow the on-screen instructions to
automatically install the device drivers. See the TSW14J57EVM user's guide (
SLWU092
) for specific
instructions.
3.7
Turn On the ADC12DJ3200EVM Power Supplies and Connect to the PC
1. Turn on the 5-V power supply to power up the EVM.
2. Connect the EVM to the PC with the mini-USB cable.
3.8
Turn On the Signal Generator RF Outputs
Turn on the RF signal output of the signal generator connected to VIN. If external clocking is used, turn on
the RF signal outputs connected to DEVCLK and LMKCLK.
3.9
Open the ADC12DJ3200EVM GUI and Program the ADC and Clocks
The Device Configuration GUI is installed separately from the HSDC Pro installation and is a stand-alone
GUI.