![Tettex MIDAS 2881 Скачать руководство пользователя страница 96](http://html1.mh-extra.com/html/tettex/midas-2881/midas-2881_operating-instructions-manual_1092946096.webp)
86
Software
TYPE
TYPE?
<string>
X X
X X Sets or returns the type of the insulation.
If you want to set the insulation type,
<string> has to be a member of the
temperature correction file. See section
“
Menu Condition”
for more information.
Otherwise the DDR Register will be set.
Example INSUL:TYPE "Liquids :Askarel"
CORRfacto
r?
X X
Returns the actual set temperature
correction factor
AMBient
TEMPeratu
re
X X
X X Sets or returns the actual ambient
temperature. The unit corresponds to
the command "TEMP:UNIT".
HUMidity
X X
X X Sets or returns the actual ambient
humidity.
7.8.4
Measurement commands
Note: All values are formatted in the same way as on the display as f.e
"12.44 kV"
for Voltages
"99.43 pF"
for Capacitances,
"0.0033"
for Dissipation, Power Factor, a.s. o.
Following units are possible:
"p" Pico
10
-12
"n" Nano
10
-9
"u" Mikro
10
-6
"m" Milli
10
-3
"k" Kilo
10
3
"M" Mega
10
6
'"G" Giga
10
9
If the system is ranging the system will answer with "--"
Command
header1
Command
header f2
or Arg.
L
S
L
A
R
S
R
A
Commentary
ACCUracy? YES
NO
X
X Returns "YES" if the accuracy of the measured signal is
enough. This value corresponds with the colour displayed
on the system. See section “Display of Measurement
Values
”
OVERflow? YES
NO
X
X Returns "YES" if a signal has an overflow detected.
URMS?
<Nx>
X
X Effective Voltage which is applied to the test object.
Answer similar as "12.434 kV"
DF?
<Nx>
X
X Dissipation factor without temperature correction.
Answer similar as "0.0043 "