Description of the “testostor 171 V1.” dialog box

P

ASSWORD

You can create a password, which will be saved in the
logger, via this button. If a password exists, it is required
before data is loaded up to the logger or readings are
read out of the logger. If an incorrect password is
entered, the function is not carried out. 

If the function is confirmed without entering characters,
the password function is deactivated.

Changes to an existing password must be entered
beforehand.

If you have forgotten what your password is, please
contact our Service Department. 

S

AVING THE PROGRAM

... 

Program settings can be stored on a data carrier (hard
disk, disks etc.) to predefine certain standard
programming for the logger. These can be opened, if
required, via 

O

PENING THE PROGRAM

... 

and transmitted to

the logger. 

O

PENING THE PROGRAM

...

You can load the programming data for the logger from
the disk via this button. In this way, loggers with
predefined standard programs can be programmed
without the need for time-consuming settings. 

S

TART

/S

TOP

This button starts and stops measurements which were
programmed in the 

Remote controlled 

starting criteria.

Every 

measurement can be stopped via the stop button.

Description of the “testostor 171 V1.” dialog box

The D

ATE

and T

IME

are adapted to the date and time of

your PC via the 

Synchronisation 

button.

You can switch between °C and °F in U

NITS

.

The control box D

ERIVATE

enables the additional

calculation of the derived values 

Dew point 

and 

Absolute

humidity 

(in g/m

3

).

H

UMIDITY CALIBRATION

Via this button, you can access a window offering you
the option of calibrating the internal or external humidity
sensor (provided you have connected Type 171-1,171-2,
1722, 171-3 or 171-6 data loggers). 

Selects from the options:

Calibration

→ 

Process to conform with the technical data of the

data logger i.e. the standard two point calibration with
11.3 or 75.3 %RH.

and 

Adjustment

→ 

Fine adjustment to optimise the calculated

absolute humidity values in a limited measuring
range by parallel relocation of the calibration
characteristic curve (by comparing and matching with
a reference instrument e.g. with reference to °C td).

Calibration is described in the “On the job...”
Chapter.

Содержание Testostor 171

Страница 1: ...testostor 1 17 71 1 Data logger and software module Instruction Manual and Software Description...

Страница 2: ...2 On the job Installation 12 Programming the logger Application information 13 Reading the memory Application information 15 Error messages 16 Humidity calibration 17 Adjustment 19 Changing batteries...

Страница 3: ...ation The testostor 171 1 171 2 1722 171 3 171 4 and 171 6 logger housings come under protection class IP 65 the testostor 171 0 logger housing comes under protection class IP 68 the testostor 171 8 l...

Страница 4: ...trument driver sets up communication between the testostor 171 series and your PC Optimum use of all the functions is only possible with the mouse i e your PC must have two free interfaces COM1 and CO...

Страница 5: ...internal 10 to 50 C X 50 to 120 C X X X external Type K TE plug in connection 200 to 1100 C X Type T TE plug in connection 50 to 350 C X Resolution 0 1 C X X X X X X X X 1 C from 250 C X Accuracy in p...

Страница 6: ...suring head 0602 0092 Sensor Thermocouple Measuring range 60 to 130 C Short term to 280 C t99 5 s Clamp probe 0602 4692 Sensor Thermocouple Measuring range 50 to 100 C Cable length 1 2 m t99 5 s Magne...

Страница 7: ...d to enter licence number see sticker on CD ROM It may be necessary to reboot when installing for the first time Note If the number entered is not accepted check if the caps key is activated if Num in...

Страница 8: ...mouse button on the testostor 171 icon and select INSTRUMENT SELECTION If communication between the PC and the data logger has been successful the window opposite appears with the status information...

Страница 9: ...ate humidity probe is included with the calibration set which you can attach to the 12 mm probe Screw the probe together with the adapter into the container the standard protective cap should remain o...

Страница 10: ...anteed that the logger sensor system and the reference instrument have the same climatic conditions the same temperature is particularly important The humidity containers are not needed for this fine...

Страница 11: ...aved measured data is lost the calibration data is saved In case of emergency please contact an authorised service point You can change the battery yourself using the battery set which is available as...

Страница 12: ...limit values are exceeded or are not reached acoustically and optically You can call up a table summarizing the number of times the limits have been exceeded by clicking twice in the limit value on t...

Страница 13: ...dings The total number of readings to be recorded per channel corresponding to your input for Start and Measuring rate are shown here Battery life The testostor 171 series is equipped with long life L...

Страница 14: ...ignal cricket sound be given off when limit values are exceeded Note the relay in testostor 1722 is activated once the limit values are determined in the probe configuration See page 23 The connected...

Страница 15: ...me consuming settings START STOP This button starts and stops measurements which were programmed in the Remote controlled starting criteria Every measurement can be stopped via the stop button Descrip...

Страница 16: ...tostor 171 2 171 3 0554 0640 Teflon sintered filter suitable for measurements in corrosive substances for 0636 9717 0554 0756 Teflon sintered filter suitable for measurements in corrosive substances t...

Страница 17: ...P Tel 19 37 31 58 00 Fax 19 37 31 58 19 testo testo com br BULGARIA Global Test OOD 1408 Sofia Tel 2 9 53 07 96 Fax 2 9 52 51 95 glbl_tst sps bg CHILE ANWO S A Santiago Tel 2 7 31 00 00 Fax 2 2 73 04...

Страница 18: ...95 20 Fax 93 753 95 26 info testo es SWEDEN Nordtec Instrument 40241 G teborg Tel 31 704 10 70 Fax 31 12 50 42 nordtec nordtec se SWITZERLAND Testo AG 8604 Volketswil Tel 1 9 08 40 50 Fax 1 9 08 40 5...

Страница 19: ...0971 2530 T Ni PC_qxd 08 02 Testo AG Postfach 11 40 D 79849 Lenzkirch Testo Stra e 1 D 79853 Lenzkirch Tel 0 76 53 6 81 0 Fax 0 76 53 6 81 1 00 E Mail info testo de http www testo de...

Отзывы: