Device components and controls
Testo Sensor LD pro V2.00
Page 13 of 48
7.9
Assembly with Gosseneck
Due to its flexibility, the gooseneck tool is used for punctual measurements in hard-to-reach areas.
Connection to the Testo Sensor LD pro is via the supplied spiral cable, see Figure 10.
It is no longer possible to use the camera.
To remove the component, remove the connection cable by pressing the release button on both sides
and pulling off the cable.
Picture 10
7.10
Assembly with Parabolic mirror
The parabolic mirror is used for measurements at greater distances as well as for high requirements
regarding selectivity and location of leakage positions.
Connection to the Testo Sensor LD pro is via the supplied spiral cable, see Figure 11.
To remove the component, remove the connection cable by pressing the release button on both sides
and pulling off the cable.
Picture 11
Note:
To use the parabolic mirror and gooseneck, these components must be activated during initial
commissioning in order to save the component-specific adjustment parameters. If this has not
already been done ex-works, the data for this is supplied via USB stick. For the activation
(parameter import), see chapter „Import measurement tool".
Содержание Sensor LD pro
Страница 43: ...Appendix Testo Sensor LD pro V2 00 Page 43 of 48...
Страница 44: ...Appendix Testo Sensor LD pro V2 00 Page 44 of 48...
Страница 45: ...Notes Testo Sensor LD pro V2 00 Page 45 of 48...
Страница 46: ...Notes Testo Sensor LD pro V2 00 Page 46 of 48...
Страница 47: ...Notes Testo Sensor LD pro V2 00 Page 47 of 48...