5 Parameterizing, adjusting and analyzing software (P2A software)
54
Field Explanation
Unit/analog output
(graphic)
Unit: 0 to 1 V/5 V/10 V or 0...20 mA, 4 to 20
mA.
Vertical: Current version of the analog output
(cannot be changed).
Horizontal: Min./max. scale end points of
selected unit.
The curve changes in accordance with the
entered value of scale minimum and
maximum.
Scale minimum/
maximum
The endpoints of the scaling can be selected
up to the stored scale minimum and
maximum. In the process, scaling can take
place beyond the measuring range in order
to adjust the analog output to the customer
system.
Unit
Selection of the physical unit.
When changing the unit, standard values are
set for scale minimum and maximum.
Signal delay
(graphic)
Curve changes according to the set signal
delay.
Signal delay
Time interval in stages 1 – 15:
1 = no delay
15 = longest delay.
The signal delay is added to the reaction time
of the sensor. The signal delay shows
averaging over the time interval of the
selected stage in seconds:
Example
Stage 10 = average of the readings from the
last 10 seconds.
The delay of the signal in relation to the change in the
process is also significantly influenced by the selection of
the particle filter.