4 Transmitter
24
During the self-adjustment phase, the analog outputs, the display
value and output value are "frozen". A defined status signal can
alternatively be chosen as output value.
The display shows
Self-adjustment active
, until the self-
adjustment phase is finished. The self-adjustment time (incl.
heating time, calculation time, cooling time) is 30 minutes. The
cycle time can be edited via the operating menu or the P2A
software, see Setting cycle time of the self-adjustment page 37 and
Self-adjustment page 59.
Deactivating the self-adjustment function of the testo 6781
will reduce measuring accuracy and should therefore be
restricted to the shortest possible length of time.
The time until the next self-adjustment is displayed in the status
bar, in intervals of approx. five seconds.