8. Maintenance
Calibrating the O
2
measuring cell
Your readings will be incorrect if full adjustment has not been
carried out. Adjustment takes 15 minutes.
If a flue gas probe is connected, it should be located in fresh
air during adjustment.
1 Switch on analyser by pressing
and press
and
simultaneously while version number and rechargeable
battery/battery capacity are being displayed (duration: 3s)
-
02-Calib
lights up
➀
.
testo 325M
:
2 Start calibraiton by pressing
.
testo 325XL
:
2 Start calibration by pressing
twice.
-
The pump starts,
O
2
and
CAL
flash and the minutes until the end
of the calibration are displayed
➁
.
-
Once calibration is complete, zeroing is carried out and is
then ready for operation.
Entering cell coefficients for CO or CO/H
2
measuring cell
The readings will be incorrect if the correct cell coefficients
are not entered! You will find the cell coefficients on the sheet
enclosed with the spare measuring cell.
If input is incorrect:
Cancel procedure with
and enter cell coefficients once
again.
1 Switch on analyser by pressing
and press
and
simultaneously while version number and rechargeable
battery/battery capacity are being displayed (duration: 3s)
-
02 Calib
lights up
➀
.
2 Select
CO Coeff
via
➁
and confirm selection with
.
-
The first digit in the first cell coefficient flashes
➂
.
3 Set digit using
or
.
4 Advance to the next positions and change to the next cell
coefficients using
. Set digits using
or
.
6 Complete input by pressing
.
-
The set values are accepted. The analyser carries out zeroing
and is now ready for operation.
➁
➀
➂
➁
➀