6 Using the product
32
5.
Set the absolute pressure with
or
and confirm with
.
If you wish to use the preset values for the pipe volume,
continue directly to the begin of the measurement
procedure as follows:
> Confirm preset value with
.
Three circular diameters (in mm) and three pipe lengths (in m) can
be entered, from which three partial volumes can be calculated.
The performance volume is calculated by adding the three partial
volumes together.
6.
Use
or
to select
Circle 1
.
7.
Confirm the selection with
.
8.
Set value using
or
. Hold the key down to go
forward/back quickly.
9.
Confirm the selection with
.
10.
To set other values (
Length 1
for first partial volume,
Circle 2
und
Length 2
for second partial volume,
Circle 3
and
Length
3
for third partial volume), perform steps 7. to 9. accordingly.
11.
Confirm
Finish
message with
.
12.
Confirm set pipe volume with
.
Begin measurement procedure
1.
Pressurize the measuring instrument.
2.
Begin measurement procedure with
.
- The stabilization phase ends. Following this, the measuring
phase begins automatically.
The stabilization phase can be ended early:
> Press
.
- The measuring phase starts automatically.
- After the measuring phase ends the difference value and the
leak rate are shown.
> Press
to save the measurement result.