![TESTO 312-4 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/testo/312-4/312-4_instruction-manual_1092247027.webp)
6 Using the product
27
6.2.4.
Save readings/implement measurement program
For the readings to be saved, the instrument must be in the
measurement view. Depending on the set save mode, the save
process is performed differently:
Save mode
Use
Manual
Retaining the current readings.
Automatic
Saving the readings for an extended
period of time.
A typical application is testing a regulator,
for example. The saved readings can be
transferred to the EasyHeat PC software
and analyzed in graph form.
Fast
Saving 25 readings per second to record
fast changes in pressure over a short
period.
A fast measurement is only possible with
pressure probes or an internal pressure
sensor. Only 1 channel can be evaluated
and the following sequence applies:
external pressure sensor before internal
pressure sensor, channel 2 before
channel 1.
The saved readings can be transferred to
the EasyHeat PC software and analyzed
in graph form.
> Before you save the readings, you must select the location
under which the data is to be saved.
Manual
save mode is set:
>
Press
to save the current readings with the date, time,
location and other available parameters.
-
flashes briefly.
Automatic
save mode is set:
>
Press
to start the set measurement program.
-
flashes for as long as the measurement program is
running.
> The save program can be ended early by pressing
. Press
again to save a new series of measurements.