background image

 

 

 

testo 512-1 and testo 512-2 

Digital differential pressure measuring instruments 

0560 1512 

0560 2512 

 

Instruction manual 

 

 

 

Содержание 0560 1512

Страница 1: ...testo 512 1 and testo 512 2 Digital differential pressure measuring instruments 0560 1512 0560 2512 Instruction manual...

Страница 2: ...e product 12 Controls on the measuring instrument 12 Implementing settings 14 Opening configuration mode 15 Setting the unit 15 Setting the area testo 512 1 only 15 Setting the Pitot tube factor P FAC...

Страница 3: ...t an application 25 Making measuring instrument settings 26 Configuring the measuring instrument menu 28 Setting Auto Off 29 Activating damping 29 Configuring alarms 30 Setting the Correction Factor 3...

Страница 4: ...the technical data Do not apply any force Do not operate the instrument if there are signs of damage to the housing or connected cables Dangers may also arise from objects to be measured or the measu...

Страница 5: ...nformation denoted by the following warnings Implement the precautionary measures specified DANGER Risk of death WARNING Indicates possible serious injury CAUTION Indicates possible minor injury CAUTI...

Страница 6: ...ng instruments testo 512 1 and testo 512 2 are compact digital pressure gauges with temperature compensation for measuring the positive and negative overpressure and the differential pressure of non a...

Страница 7: ...iption 6 Product description Instrument overview testo 512 1 1 Control keys 2 Display 3 Connections for pressure hoses 4 Speaker for alarm sound 5 Battery compartment Explanation of icons Refer to ins...

Страница 8: ...5 Product description 7 testo 512 2 1 Control keys 2 Display 3 Connections for pressure hoses 4 Speaker for alarm sound 5 Battery compartment Explanation of icons Refer to instruction manual...

Страница 9: ...nsert or replace batteries 3 x AA batteries Observe the polarity 3 Close the battery compartment When not in use for a long period Take out the rechargeable batteries Symbol explanation Do not allow c...

Страница 10: ...itching on 1 Press and hold down 2 sec the On Off key Measurement view is opened The current reading is displayed or lights up if no reading is available Switching off 1 Press and hold down 2 sec the...

Страница 11: ...tooth is enabled for approx 60 seconds to establish a connection to the testo Smart App While the instrument is trying to establish a Bluetooth connection the icon flashes on the display If the app co...

Страница 12: ...quires Bluetooth 4 2 Bluetooth is enabled in the measuring instrument 1 Open the testo Smart App The app automatically searches for Bluetooth devices in the vicinity 2 In the Sensors menu check whethe...

Страница 13: ...h Settings and controls are implemented either on the instrument or via the app If the measuring instrument is connected to the testo Smart App settings can only be made via the app The measuring inst...

Страница 14: ...ty reading 7 Parameter 2 selected unit 8 Print key 9 Illumination MENU ENTER key testo 512 2 1 On Off MODE END key 2 0 key 3 Battery indicator 4 Differential pressure reading 5 Parameter 1 unit 6 Prin...

Страница 15: ...settings all settings are transferred immediately There is no Cancel function Function Setting options comments Carry out zero point adjustment long press Carry out zero point adjustment Arrow left Fr...

Страница 16: ...The instrument is now in configuration mode Press MENU ENTER to switch to the next function You can leave configuration mode at any time To do this press MODE END until the instrument has changed to...

Страница 17: ...he measuring instrument displays T_AMB and the ambient temperature last measured The ambient temperature is required for internal calculations in the measuring instrument and is measured but not displ...

Страница 18: ...enable disable the alarm sound ON OFF and confirm with MENU ENTER Enabling Bluetooth Configuration mode is open is displayed 1 Press to enable disable Bluetooth ON OFF and confirm with MENU ENTER Carr...

Страница 19: ...easurement o Differential pressure measurement Connecting a pitot tube testo 512 1 only 1 Connect the pressure hoses 4 or 6 mm to the pressure connection nipples with the correct signs 2 Attach the pr...

Страница 20: ...m minimum value The current reading can be frozen The maximum and minimum values since the last time the instrument was switched on in the standard view or during a multi point or continuous measureme...

Страница 21: ...reading is displayed in the lower line 2 To record readings in the desired quantity Press MENU ENTER several times 3 To end measurement and calculate the mean value Press MODE END and flash The number...

Страница 22: ...e the mean value Press MODE END and flash The measurement period and the calculated continuous mean value are displayed 5 To switch back to the measuring view Press MODE END Printing data A testo Blue...

Страница 23: ...view and settings menus cannot be opened on the measuring instrument Overview of operating controls 1 Choice of applications 2 Switch between the views live graphic table 3 Display of connected measu...

Страница 24: ...nt data report Searching Favourite Deleting Further information Display report Multiple selection The Main menu can be accessed via the icon at top left To exit the main menu select a menu or right cl...

Страница 25: ...you can find the version number of the installed App 1 Click on 2 Select Help and Information 3 Select Instrument information The version number of the app and the ID are displayed Displaying the tuto...

Страница 26: ...Your selected application is displayed Setting favourites 1 Click on A selection of menus for various applications is displayed 2 Select the application you want to set as a favourite 3 Click on The...

Страница 27: ...ensors The Sensors menu opens 3 Click on the required measuring instrument Information about the model order number serial number and firmware version is displayed 4 Click on the Settings tab A window...

Страница 28: ...pen an input window to enter a specific value or select a unit For details on the settings options see the following sub sections Changes to the measuring instrument settings in the app are transferre...

Страница 29: ...tab in the Sensors menu is open 1 Activate Customization instrument menu and click on the blue text Change instrument menu list under the settings heading The Customization instrument menu dialogue op...

Страница 30: ...the menu reset M RES and then all settings menus will be displayed on the measuring instrument again Setting Auto Off The Settings tab is open 1 Enable Activate Auto off using the slider The measuring...

Страница 31: ...dow for activating and defining upper and lower warning and alarm values is displayed 5 Click OK to confirm the settings Changes to the measuring instrument settings in the app are transferred directl...

Страница 32: ...e value for the cross sectional area and confirm with OK The changes are transferred to the measuring instrument and synchronization with the app is confirmed on the measuring instrument via SYNC DONE...

Страница 33: ...NE Display of the readings The available readings can be displayed in different views Live view The readings transmitted by the measuring probes can be displayed in a live view Readings from all conne...

Страница 34: ...hide a measuring instrument s measurement channel 4 Click to select the unit of a measurement channel 5 Click OK to confirm the settings Exporting readings Export Report Delete 1 Click on 2 Select Me...

Страница 35: ...all readings button 3 Click on Create For measurements please be aware that the option Create PDF with all readings is only possible up to 30 pages due to the resulting file size and number of pages...

Страница 36: ...ap lock 2 Insert or replace batteries 3 x AA batteries Observe the polarity 3 Close the battery compartment When not in use for a long period Take out the rechargeable batteries Cleaning the instrumen...

Страница 37: ...Pa 20 1 200 hPa Measuring range 0 to 200 hPa Operating temperature 20 to 50 C Storage temperature 20 to 50 C Operating humidity 0 to 80 RH IP class IP20 Level of contamination PD2 Max operating altitu...

Страница 38: ...Measuring range 0 to 2000 hPa Operating temperature 20 to 50 C Storage temperature 20 to 50 C Operating humidity 0 to 80 RH IP class IP20 Level of contamination PD2 Max operating altitude 2000 m abov...

Страница 39: ...play BT Fail Bluetooth connection could not be established Check Bluetooth connections Restart measuring instrument restart testo Smart App Display Print Fail Printout could not be performed successfu...

Страница 40: ...Accessories and spare parts Description Order no Bluetooth IRDA printer 0554 0621 For a complete list of all accessories and spare parts please refer to the product catalogues and brochures or visit...

Страница 41: ...Testo SE Co KGaA Celsiusstr 2 79822 Titisee Neustadt Germany Phone 49 0 7653 681 0 E mail info testo de www testo com 0970 9512 en 01 10 2022...