5. Technical data

Feature

Values

Sound pressure level

94dB/114dB 
in reference conditions,
switchable

Accuracy: 

Sound pressure level

±0.5dB

Frequency

1000Hz ±2%

Reference conditions
- Temperature

+23°C (74°F)

- Relative humidity

50%

- Air pressure

1013hPa

Operating conditions
- Temperature

0 to +40°C (32 to 140°F)

- Relative humidity

10 to 90%RH

- Air pressure

650 to 1080hPa 
(to 2000m amsl)

Dependency coefficients
- Pressure dependency

0.005dB/hPa

- Temperature dependency

0.005dB/°C

- Humidity dependency

0.005dB/%RH

Distortion factor

<3%

Storage and transport
conditions
- Temperature

-10 to 50°C (14 to 122°F)

- Relative humidity

0 to 70%RH

Battery

9V block (6F22)

Battery life

Approx. 40 hours 

en

Содержание 0554 0452

Страница 1: ...Instruction manual en testo sound level calibrator 0554 0452...

Страница 2: ...e product before putting it to use Keep this document to hand so that you can refer to it when necessary 2 Product Description Function and battery status display LED On Off and level switch Battery c...

Страница 3: ...ithin the parameters specified Do not use force Do not expose to electromagnetic radiation e g microwaves induction heating static charge heat or strong fluctuations in temperature Never store with so...

Страница 4: ...osal Dispose of defective rechargeable batteries and spent batteries at the collection points provided Send the product directly to us at the end of its life cycle We will ensure that it is disposed o...

Страница 5: ...ure 1013hPa Operating conditions Temperature 0 to 40 C 32 to 140 F Relative humidity 10 to 90 RH Air pressure 650 to 1080hPa to 2000m amsl Dependency coefficients Pressure dependency 0 005dB hPa Tempe...

Страница 6: ...ed voltage is then too low Dimensions in mm lxbxh 113 x 63 x 44 Weight Approx 170g incl battery EC guideline 89 336 EEC 6 Initial Operation 1 Open battery compartment 2 Insert battery observe 3 Close...

Страница 7: ...microphone 2 Switch on instrument and set the measurement range from 80 to 130 dB 3 Set time weighting Fast Slow and frequency weighting dB A dB C to the values in which the instrument is usually oper...

Страница 8: ...000 852 795 0 3 Example Height above mean sea level 500 Target value 94dB 0 1dB 93 9 dB If the value displayed deviates from the target value the instrument should be recalibrated Using a small screwd...

Страница 9: ...d by recalibration please contact your dealer or Testo Customer Service You will find contact details at www testo com 8 Care and Maintenance Changing the battery If the battery voltage needed to oper...

Страница 10: ...22 9 Accessories Name Part no Calibrator 0554 0452 incl battery case and inch adapter 9V rechargeable battery 0515 0025 Recharger 0554 0025 for external recharging of rechargeable batteries...

Страница 11: ...23 en...

Страница 12: ...testo AG Postfach 11 40 79849 Lenzkirch Testo Stra e 1 79853 Lenzkirch Telefon 07653 681 0 Fax 07653 681 100 E Mail info testo de Internet http www testo com 0973 0800 01 C wh 20 07 2004...

Отзывы: