Operation manual Pneumator
∙
Testo Industrial Services
∙
Version 1.4
∙
2019
∙
Page 25
12.
Technical data
The resolution of the 4 Pneumator models is as follows:
:
Model 1hPa: 100,01 Pa -> Resolution 0,01
Model 10 hPa:
10,001 hPa -> Resolution 0,001
Model 100hPa:
100,01 hPa -> Resolution 0,01
Model 1000hPa
1000,1 hPa-> Resolution 0,1
13.
Trouble-shooting
Interferance
Reason
Procedure
Device does not function,
no supply voltage
Check whether the plug is connected
display is dark
right in the IEC socket
Switch on the device with the main
switch on the rear side
Check the fuse and if necessary, replace
the fuse with a new one
Attention! Pull the power plug
Set pressure is not being
leak source
Slide on the hoses correctly, eliminate
reached,
in the system, too big possible leaks
pump runs permanently
hose diameter
hose diameter max 5mm
20..26 VDC