§3

 

Preparation 

 

§3

 

Preparation   

 
3-1

 

Preparation to be executed first after installation   
When this tester  is used for the  first time after  mounting  (or relocating) it, perform the following 
preliminary operation and confirm its operation according to the following procedure.   

 

1)

 

Power supply   
A single-phase 220VAC power supply is necessary as the control power supply for this tester.    An 
earth  plug  is  mounted  at  the  tip  of  the  power  cord  of  the  control  unit  Connect  the  earth  to  the 
ground without fails.   

 

2)

 

Pneumatic pressure source   
Pneumatic  pen  cylinders  are  used  for  fixing  a  specimen.    Feed  the  air  pressure  from  the 
compressor

2

  (option) or indoor piping (higher than 0.49MPa) in a building.   

The  air  pressure  supplied  to  the  tester  body  is  regulated  by  a  filter  regulator.    The  regulated  air 
pressure operates the pneumatic pen cylinders through the solenoid valves. Since the air pressure 
system  is  operated  by  feeding  power  supply,  the  pneumatic  pen  cylinders  are  controlled  to  fix  a 
specimen so long as the air pressure is supplied when turning off the power supply.   

 

                                                

2

 

If the compressor (option) is provided with a start/stop (ON/OFF) switch, turn on the switch before starting the test.

 

Содержание DT-202

Страница 1: ...DT 202_007 Operation Manual for Drop Tester Model DT 202 SHINYEI TESTING MACHINERY CO LTD...

Страница 2: ......

Страница 3: ...ence of this display during work may cause personnel death or serious injuries Observe the instruction contents Negligence of this display during work may cause a fire an electric shock accident or ot...

Страница 4: ...the control unit from the case during test otherwise it may cause ingress of foreign substances to result in a trouble It is possible that parts are replaced and checked due to unavoidable circumstan...

Страница 5: ...tester base interior 6 A specimen may tilt down in an unexpected direction due to drop collision Particularly be careful since certain specimens may be broken to scatter 7 If the cover is removed from...

Страница 6: ...X 81 78 332 1619 NAGOYA OFFICE Nishiki emuai BLDG 5F 1 7 39 Nishiki Nakaku Nagoya Aichi 460 0003 Japan Tel 81 52 218 5472 Fax 81 52 218 6112 For inquiries through telephone Monday to Friday Excluding...

Страница 7: ...5 4 4 Specimen supporting base 5 4 5 Height setting pointer 5 4 6 Pneumatic pen cylinders for fixing and releasing a specimen 5 4 7 Contacts 5 4 8 Pneumatic regulator 6 5 Operation method 7 5 1 Contro...

Страница 8: ......

Страница 9: ...is designed by taking a serious view of safety 1 When the pneumatic cylinders are set to the pull side the drop operation is not done even if DROP push button switch is pressed 2 Drop operation is do...

Страница 10: ...2 Outer View 2 2 Outer View Fig 1...

Страница 11: ...ithout fails 2 Pneumatic pressure source Pneumatic pen cylinders are used for fixing a specimen Feed the air pressure from the compressor2 option or indoor piping higher than 0 49MPa in a building The...

Страница 12: ...for fixing a specimen in particular for normal condition before starting the test See Fig 2 3 Drop height setting Set the height setting pointer to an optional position of the scale and fix it by fast...

Страница 13: ...specimen supporting base is provided with two pneumatic pen cylinders for fixing and releasing a specimen The drop height is set to the center axis of these pneumatic pen cylinders 4 5 Height setting...

Страница 14: ...ders for fixing and releasing a specimen Please set it over 0 49MPa at least Fig 4 2 SET adjusting speed controller This speed controller is mounted on the pneumatic pen cylinders This airflow regulat...

Страница 15: ...pneumatic pen cylinders are controlled to fix the specimen when the power supply has been turned off 2 DROP push button switch This push button switch is used for dropping a specimen A specimen must...

Страница 16: ...5 Operation method 8 Fig 6 Control unit...

Страница 17: ...ecimen Memo If the drop height is lower than 500mm the releasing operation does not function even if the switch is set to DH 500mm Set the switch to DH 500mm without fail Memo An automatic drop height...

Страница 18: ...2 Check if the metal connector Co 2 on the rear panel of the control unit is connected See 9 1 Wiring diagram 6 3 Pneumatic pen cylinders don t operate 1 Check if the metal connector Co 1 on the rear...

Страница 19: ...ISO VG10 may cause dust to be attached easily more or less Wipe off the guide rail by a clean cloth without fail after use 7 2 The operation may change due to dust etc after a long time use Check the...

Страница 20: ...nsions 600mm in width 500mm in depth 2382mm in height 8 Tester weight approx 110kg 9 Painting color Body Munsell code 10B5 10 Japan Painting Industrial Association B69 50T Control unit Munsell code N9...

Страница 21: ...8 Performance and specifications 13 8 4 Attachments 1 Single ended wrench 8mm across flats 1 pc 2 Test data 1 copy 3 Operation manual 1 copy Fig 7 Maximum specimen size...

Страница 22: ...9 Diagram 14 9 Diagram 9 1 Wiring diagram...

Страница 23: ...e filter S 5 Manual switch M 2011 Nikkai 1 Single pole double throw Guard cover for Drop switch AL KH6 IDEC 1 PL 1 Power lamp S1 1 DC DC Converter CCK 2405SF TDK 1 24VDC 5VDC RC Spark killer XE1201 3...

Страница 24: ...9 Diagram 16 9 3 Pneumatic piping diagram...

Отзывы: