Chapter 7 – Maintenance
Page 7-10
TestEquity TEC1 Thermoelectric Temperature Chamber
EZ-Zone Custom Menu Parameters
Factory Page access from Home Page:
Press both the Advance
‰
and Infinity
ˆ
keys for six
seconds.
Factory Page\Custom Menu
Parameter
Prompt
Value
Custom 1
AC.Pu
Active Process Value
Custom 2
AC.SP
Active Set Point
Custom 3
h.Pr
Heat Power
Custom 4
C.Pr
Cool Power
Custom 5
AUt
Autotune
Custom 6
idle
Idle Set Point
Custom 7
P.Str
Profile Start
Custom 8
P.ACr
Profile Action Request
Custom 9 to 20
nonE
none
EZ-Zone Operations Page Parameters
Operations page access from Home Page:
Press both the Up
¿
and Down
¯
keys for three
seconds.
[``Ai]
will appear in the upper display and
[oPEr]
will appear in the lower display.
Operations Page\Analog Input 1
Parameter
Prompt
Value
Process Value
Ain
current temperature
Error Status
I.Er
current status
Calibration Offset
i.CA
0.0 °C (offset as required to meet accuracy spec)
Operations Page\Loop Menu
Parameter
Prompt
Value
Control Mode
C.M
Auto
Autotune Set Point
A.tSP
90.0 %
Autotune Request
AUt
no
Closed Loop Set Point
C.SP
23.0 °C
Idle Set Point
id.S
23.0 °C
Heat Proportional Band
h.Pb
7.0
Cool Proportional Band
C.Pb
4.0
Time Integral
ti
180
Time Derivative
td
30
Dead Band
db
0.0
Open Loop Set Point
o.SP