Chapter 7 – Maintenance
TestEquity TEC1 Thermoelectric Temperature Chamber
Page 7-7
EZ-Zone Temperature Controller Setup Parameters
!
CAUTION:
The “EZ-Zone PM User’s Manual” is a general manual and is written by the
manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
PM6R1CC-2AAAAAA, as described on page 112 of the “EZ-Zone PM
User’s Manual” are applicable. The “Retransmit” function is available as an
option.
!
CAUTION:
The EZ-Zone Temperature Controller has been properly configured by
TestEquity to match the chamber’s system requirements and to perform
optimally over a wide range of operating conditions. Improper modifications
to these setup values can result in erratic performance and unreliable
operation. Do not attempt to modify the setup values, unless you thoroughly
understand what you are doing. If there is any doubt, please call TestEquity
before proceeding.
!
CAUTION:
The EZ-Zone Temperature Controller “Alarm” function is NOT used in the
chamber and is NOT connected.
!
CAUTION:
NEVER select “Factory Default” in the Diagnostic Menu. “Factory Defaults”
are NOT the TestEquity configuration parameters for this chamber. If you
select “Factory Defaults”, you must reconfigure all System and Operation
Parameters as documented in the TestEquity manual, NOT the “EZ-Zone PM
User’s Manual”.
TestEquity has configured the Temperature Controller with the parameters as documented on the
following pages. These parameters have been saved in User Save Set 1 and User Save Set 2
which is accessible from the Diagnostics Menu.