Chapter 6 – Specifications
TestEquity TEC1 Thermoelectric Temperature Chamber
Page 6-1
Chapter 6 – Specifications
Model TEC1 Chamber Specifications
Temperature Range
+10°C to +60°C
Control Tolerance
Better than ±0.1°C
(Measured at the control sensor after stabilization)
Uniformity
±1.0°C
(Variations throughout the chamber after stabilization, at least 2-inches from any inside surface)
System Accuracy
±2.0°C
(Indication on temperature controller compared to measurement in the center of the chamber, in the range of +15°C
to +55°C, after 30 minute stabilization at set point)
Cool Down Time
Start Temp to
+55°C
+37°C
+23°C
+15°C
+10°C
+55°C
-----
7 min
16 min
26 min Ultimate
+23°C
-----
-----
-----
7 min Ultimate
Heat Up Time
Start Temp to
+23°C
+37°C
+55°C
+60°C
+23°C
----- 3.5 min
10 min Ultimate
+15°C
2.0 min
5 min
11 min Ultimate
*Note:
Heat and cool transition times are measured in an empty chamber after a 30 minute soak at the start
temperature, as indicated on the temperature controller.
Transition times do not include the effect of proportional
band when approaching the temperature set point.
Live Load Rating
+37°C
+23°C
+15°C
+10°C
90 Watts 50 Watts 35 Watts 20 Watts
Power Requirements
Input Voltage
88 to 132VAC / 176 to 264VAC, 50/60 Hz (internal switch for voltage range)
NOTE:
Model TEC1-001 is rated for 115 V nominal input only.
Current Draw
6A at 115 VAC, 3.5A at 230 VAC
Workspace Dimensions
16" W x 8" H x 13" D (0.96 cubic feet)
Outside Dimensions
24.7" W x 12" H x 20.4" D (nominal)
Min. Installed Clearance
4" from the rear
Cable Notch
1" x 3" Cable Notch on left and right side (two total).
Weight
52 pounds
Sound Level
Less than 50 dBA
(A-weighted, measured 36" from the front surface in a free-standing environment)
NOTE:
Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. This
product is designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result
in decreased cooling performance. Operation above 27°C (80°F) will have a significant impact on low-temperature
operation.