Chapter 7 – Specifications
TestEquity 1016C-1027C Temperature Chamber
Page 7-1
Chapter 7 – Specifications
Model 1016C Chamber Specifications
Temperature Range
–73°C to +175°C
Control Tolerance
±0.5°C (±0.2°C Typical) (Measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)
Live Load Capacity @
+23°C
0°C
–20°C
–40°C
–55°C
–65°C
2900 W 2600 W 2300 W 1750 W 1450 W 1050 W
Cool Down Transition Time (empty chamber)*
End Temp
Start Temp to
+23°C
0°C
-20°C
-40°C
-55°C
-65°C
+23°C
-----
1 min
5 min
10 min
15 min
20 min
+85°C
7 min
12 min
17 min
23 min
29 min
35 min
Heat Up Transition Time (empty chamber)*
-40°Cto +85°C
15 min
*Note:
Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber,
at the control sensor as indicated on the temperature controller, at a 23°C ambient. Measured with the temperature
controller set beyond the end temperatures. Does not include the effect of proportional band when approaching the
temperature set point.
To calculate rate of change for a particular condition, take the difference between the Start
Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty):
From +85°C to -40°C = 125°C / 23 min = 5.4°C/min.
Heat Up Example (empty):
From -40°C to +85°C = 125°C / 15 min = 8.3°C/min.
Input Power Requirements
230 V ±10%, 60 Hz, 3 PH
Max Current Draw 39 A; Recommended Service 50 A
208 V -5/+10%, 60 Hz, 3 PH
Max Current Draw 35 A; Recommended Service 45 A
Input may be configured for nominal 208 V or 230 V in the field by changing jumpers.
Call for other voltages or 50 Hz operation.
Workspace Dimensions
30" W x 30" H x 30" D (16 cubic feet)
Outside Dimensions
38" W x 78.5" H x 56" D (nominal)
Door latch adds 3" to width on right side.
Min. Installed Clearance
12" from the left and right side, 24" from the rear
Access Ports
4" Port on left and right side (two total), Supplied with foam plugs
Sound Level
68 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from
the floor, in a free-standing environment)
Heat of Rejection
27,500 BTUH (maximum rated chamber load at maximum cooling rate from high
temperature soak)
Weight
1130 pounds
NOTE:
Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. This
product is designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result
in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C
(85°F) or below 16°C (60°F) ambient is not recommended.