Chapter 3 – Operation
Page 3-8
TestEquity 1007H Temperature Chamber
Relief Vent
Excess pressure in the chamber workspace is vented through a pressure-relief check-valve,
which is located on the top of the chamber behind the fan shroud. There is no need to vent this
externally.