RS-232C Commands
33
5.6 Error Data List
Table 5-3
List of Error Data
Error Code
Description
ERR10
Syntax error. The command does not exist or is invalid.
ERR40
Status Error, It occurs when an invalid command is sent. For example, Probe
UPWARD command while the lid is closed. It is returned if the command is sent
when the motion can not be done in conjunction with the current status.
ERR50
Lid Open Sensor error. It occurs only to certain products which should always detect
whether the lid is open. The response comes when the sensor does not detect open
lid.
ERR60
Safety error. It is returned when a problem in multiple motions occurs.
ERR90
It occurs when the command is sent while the Probe is in Off Mode.
5.7 RS-232C Commands
5.7.1 System Commands
MODEL?
COUNT?
MODEL?
A query to see the model name, firmware version, and date of manufacture.
Type
Query
Input
MODEL?
Response
<Name>_<Version> <Date>
Example
MODEL?
TC-5910DP_STD_Ver: 5.2 2019.04.09