TESA-hite 400 / 700
4.14 Measuring perpendicularity deviations
Perpendicularity deviations and run-out errors are recorded using a lever indicator (e.g. TESATAST).
This is fitted to the mounting pin (N° 00760222), which itself will be fixed to the probe-fixing arm of
the TESA-Hite 400 / 700. When measuring perpendicularity, it is recommended that you clamp the
measuring head.
Clamping of the measuring head: Move the measuring slide upward until the stop; exert a force on
the handwheel, like if you would do a probing up; this will lock the transport carriage on the
measuring carriage. To unlock the carriage do the same, but in the opposite direction.
The measuring is effected by moving the measuring head vertically while observing the maximum and
minimum deviations recorded by the lever indicator. When measuring run-out errors, it is also
necessary to tighten the knurled screw locking the measuring-head drive 10.
It is also possible to measure the perpendicularity deviations using a digital instrument. To do it, fix
the digital instrument on the probe fixing arm (N° 00760222) and connect the output on the interface
(N° 04760070), itself connected on the RS connector of the TESA-Hite. It is recommended to use a
GT21 inductive probe connected to a TT20. The bidirectional cable (N° 04761049) has to be used.
The measuring time is 70ms that means 14 measurements per second. By using a DIOGICO 12 the
measuring time will be 150ms that means 6 measurements per second.
L
min.
10 mm
.../L
A
A
36
Содержание TESA-hite 400
Страница 1: ...TESA hite 400 700 Instruction manual TESA hite 400 700 ...
Страница 2: ...TESA hite 400 700 2 ...
Страница 30: ...TESA hite 400 700 2 x mm 2 1 x 5 x mm 3 1 x mm 4 2 x 1 x mm 5 1 x 1 x mm 6 4 x 1 x mm 7 5 x 1 x mm 8 30 ...
Страница 32: ...TESA hite 400 700 mm mm mm PRESET function Mode 3 Access and entry of PRESET As for Modes 1 and 2 32 ...
Страница 52: ...TESA hite 400 700 52 NOTES ...