Detailed procedure for detecting the culmination point
Probe/Contact
Display
Active keys
Comments
mm /in
Recording constant OK.
Probe reference
mm
mm
Reference OK.
mm
Probe diameter
Internal, downward.
Normal probing
Beep
mm
• Place the gauge head slightly off-
centre with respect to the bore axis.
Wait 1 second
• After waiting about 1 second in the
In probing zone
probing zone.
•
Culmination
proposal:
if
passing
through the neutral point without
culmination=normal
probing.
• Adjust the force in upper zone
mm
Move the part or the TESA-Hite 400/700
so that the gauge head passes through the
maximum or the minimum point.
Culmination point detected.
Beep
The bar graph indicates the current
measuring force.
23