Instruction manual for MH & MH+M
78
5
In this precise case, the software expects to detect
a minimum internal culmination point. If the
workpiece is moved in a way, that the probe is
pulled away from the culmination point, the LED
will warn you about the incorrect handling.
6
The LED became red at step 5. Therefore, it is
important to move the workpiece in the opposite
direction in order to detect the inflection point. This
is the correct direction and the LED becomes
orange.
7
As long as the probe has not passed the inflection
point and is not “going up” on the other side of the
point to be measured, the software understands
that the minimum point has not yet been
determined. Therefore, it is important to continue
to move the workpiece until the LED becomes red
(which is equivalent to saying that the probe is
moving away from the culmination point).
8
Change the direction again. Carefully, you are
approaching the zone of the culmination point.
9
As soon as you have reached the culmination
point, the LED becomes green. Make sure you are
at the right place by verifying the values displayed
on the screen.
You can then unlock the ring and measure the
culmination point.
10
-
If you continue to move the measuring probe in any
direction from the culmination point, the LED will
become red again.
12.11 Bore
measurement,
static mode (MH)
The measurement of a bore or axis follows the same steps as the procedure described
here. The only difference is that the determination of the bore or axis diameter requires a
probing up and a probing down instead of a single probing down as described in step 10.
Thus, mode ST2
is required as probing with two contacts
up and down is required.
12.12 Dynamic
culmination point
(MH)
On the manual MICRO-HITE, a culmination point can be measured in dynamic mode
meaning by moving the workpiece back and forth, so that the probe passes the maximum
or minimum culmination point to be detected. The height of the culmination point is then
measured on the fly and stored in the memory. Measurements in dynamic mode are only
possible when the corresponding option has been configured in the system parameters
(for further details, see this chapter).
At each passage, a new culmination point is calculated and compared
with the previous ones. If the difference between all the memorised
points is above a certain limit, the measurement is considered as invalid.
Process
1. Place the probe inside the bore.
Содержание MICRO-HITE
Страница 10: ...Instruction manual for MH MH M 10 Fig Description of the constitutive elements of the TESA MICRO HITE M ...
Страница 83: ...Instruction manual for MH MH M 83 ...
Страница 99: ...Instruction manual for MH MH M 99 ...
Страница 152: ...Instruction manual for MH MH M 152 ...
Страница 163: ...Instruction manual for MH MH M 163 EXAMPLE TESA WORKPIECE ...