Instruction manual for MH & MH+M
81
that will move and the inflection point is determined on the fly. Thus, the
static mode is not available for this type of model.
Process
1. Place the probe inside the bore.
2. Displace the probe slightly to one side of the culmination point (visually).
3. Select an action key on the keyboard
,
,
or
. The height gauge will
move in the required direction in order to establish contact with the workpiece to be
measured.
4. Once the contact is established, it is important to wait for the validation of the software
in order to proceed with the displacement of the probe in the bore/axis to be measured.
5. Once the minimum (or maximum) point passed, the software will issue a beep. The
probe will retract (displacement defined in the system parameters) and stop.
6. The result is automatically displayed on the screen of the control panel.
Содержание MICRO-HITE
Страница 10: ...Instruction manual for MH MH M 10 Fig Description of the constitutive elements of the TESA MICRO HITE M ...
Страница 83: ...Instruction manual for MH MH M 83 ...
Страница 99: ...Instruction manual for MH MH M 99 ...
Страница 152: ...Instruction manual for MH MH M 152 ...
Страница 163: ...Instruction manual for MH MH M 163 EXAMPLE TESA WORKPIECE ...