background image

 

4. C a r r y i n g   o u t   P r o c e d u r e s  

Perform the stages as indicated in Articles 1 and 2, paragraph “

Work 

sequence

”.

 

Make sure that 0 mA LED is highlighted on the indicator “

Current mode

 

(6),

 

whereas

 

indicator “

Battery

” (5) is not highlighted or blinking. 

Connect a headband with electrodes to the Unit. 
Make sure that skin at the site of fixing electrodes is clean, free of make-up, 

without rash or damage. 

A patient should remove any clips or earrings out of ears.  

Put a headband with electrodes over a patient’s head (Figure 4). For this:  

– put a headband on (15), slightly tight, avoiding an excessive 

unpleasant pressure or discomfort, by fixing it with a Velcro closure 
above forehead; 

– large frontal electrode (16) should be placed up beneath a 

headband, right in the middle of the forehead, keeping it free of hairs. 
Electrode should be in equal contact with patient’s skin; 

– small retromastoid electrodes (17) should be placed beneath a 

headband behind ears over the area of mastoid processes. Keep it out 
of hairs and wires. Make sure an electrode is in equal contact with 
patient’s skin. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Figure 4 – Position of headband with electrodes over patient’s head 

a) b) 

c) d) 

e) 

 

10

 

MEDICAL INSTRUCTION ON TRANSCRANIAL PULSED 
ELECTRIC STIMULATOR  «DOCTOR TES-03”

 

Finding out an appropriate magnitude of stimulating current

 

Press button (3), and increase the current until patient started to feel 

tingling or slight vibration beneath electrodes, or subjective feeling of light 
flickering behind closed eyes. While adapting to the applied current patient 
may lose such feelings. In this case magnitude of current may be slightly 
increased until the described feelings come back. In case the applied 
current causes an irritation, discomfort or no adaptation at all, magnitude 
of current should be reduced. For this, press button (2) until the 
uncomfortable sensation is disappeared. 

It is recommended that during the 1

st

 session stimulating current should 

not be higher than 0.5 mA. For further procedures it should not exceed 
higher than 1.5 mA (see pages 16–19). 

After procedure completed (in 30 minutes) the Unit will automatically 

reduce current reaching 0 mA, and sound signal will be heard.  

Take the headband with electrodes off of patient’s head. Disconnect the 

electrodes from the Unit.  

Switch off the Unit by pressing button (1). 

After procedure electrodes should be wiped with a clean swab soaked in 

3% hydrogen peroxide solution or alcohol. 

ATTENTION! 

In case protection mode was initiated during the procedure (Indicator 

Protection

” (4) is highlighted), intensity of stimulating current slowly 

decreases up to 0 mA as displayed on indicator “

Current Mode

” (6), go 

directly to the Paragraph “

Common Troubles And Remedies

”, page 23. 

General recommendations 

It is recommended to perform TES-therapy procedures when patient is 

in comfortable supine or sitting position, in calm atmosphere, in about 1-2 
hours after meal. 

First session is supposed to be an introductory, and aimed at adapting a 

patient to a procedure. Thus, intensity of stimulating current must be used 
at minimum, ranging up to 0.5 mA, for 15–20 min (stop time manually by 
slowly decreasing current till 0 mA). Even though patient may not have any 
subjective feelings current intensity should not be higher than recommended.  

The main criteria as to how correctly find out an individual regimen for 

each patient are tolerability as well as positive clinical effect. In vast majority 
the appropriate current intensity is supposed to be reached if patient started 
to feel tingling or slight vibration beneath the electrodes, or subjective feeling 

Содержание DOCTOR TES-03

Страница 1: ...ere awarded with Grand Prix in the nomination High Impact Invention Ten thousand professional TES Units have been successfully applied both in Russia and abroad Aside from mentioned prizes they have won many other Russian and foreign top awards V P L e b e d e v D M S c P r of e s s o r Rus Acad Natur Sci Academician laureate of the state USSR Prize laureate of the Government of the RF Prize Honou...

Страница 2: ...to fail which is followed by appearance of acute and chronic diseases pain syndromes disturbed healing processes compromised immunity psychosomatic disturbances which influence a quality of dreams performance and quality of life as a whole 4 What complaints symptoms and conditions might be associated with an endorphin deficiency Acute headache of unreasonably high intensity Unexplained repeated he...

Страница 3: ...cco and alcohol addiction normalization of the impaired stool passage SAFETY PRECAUTIONS Before reading the User s Guide it is forbidden to use the Unit Use the Unit strictly as intended Follow the work sequence as it is described in Paragraphs Work sequence and Carrying out Procedure In case a nonrepairable failure was found which is not mentioned in the Paragraph Common Troubles And Remedies con...

Страница 4: ...ee Figure 3 socket of the headband electrodes 19 to the hole at the upper end face 8 in the Unit Plug male connectors 18 in the adhesive electrodes 16 17 as shown in the Figure 3 Frontal electrode 16 has the size 40х80 mm and retromastoid electrodes 40х40 mm After that close the frontal electrode 16 onto any of retromastoid electrodes 17 by holding them with flat surfaces 15 Press button 3 and hol...

Страница 5: ...irritation discomfort or no adaptation at all magnitude of current should be reduced For this press button 2 until the uncomfortable sensation is disappeared It is recommended that during the 1st session stimulating current should not be higher than 0 5 mA For further procedures it should not exceed higher than 1 5 mA see pages 16 19 After procedure completed in 30 minutes the Unit will automatica...

Страница 6: ...cause of identical mode of action it is useless to simultaneously apply TES therapy together with acupuncture Acupuncture may be applied before or after TES sessions as an additional therapeutic approach to strengthen effect from treatment TES therapy can be well combined with psychotherapy Moreover both Of them can compliment each other and potentiate therapeutic effect 12 Due to this additive ef...

Страница 7: ...licit habituation addiction or dependency During TES treatment after the first sessions most patients start to feel improved overall condition sleeping mood Pain sense is weakened or fully vanished healing of damaged tissues is accelerated However if during TES therapy sessions patient started to feel a subjective deterioration of well being the procedure should be stopped After that there must be...

Страница 8: ...f post traumatic stress Stammering Insomnia performance decrement in healthy people 10 12 sessions per course once a day or every other day for 30 min Current up to 1 5m A R e p e a t c o u r s e i n 3 4 months 16 Cardiology Hypertension disease stage I II Hypotension syndromes Neurocirculatory vegetative vascular dystonia Patients recovering after uncomplicated heart attack 10 12 sessions per cou...

Страница 9: ...up to 1 5 mA Repeat course according to indications it may correspond as close as possible to the previous one 18 Obstetrics and Gynecology Menstrual disorders in pubertal girl Premenstrual syndrome Menstrual disorders and endocrine infertility under obligatory medical supervision Vomiting during recent pregnancy Menopausal disorders 10 15 sessions per course once a day for 30 min Current up to 1 ...

Страница 10: ... the Unit can be cleaned by using 70 alcohol A headband may be washed depending on frequency of use For this headband should be disconnected from the wire that plugs it into the Unit as well as from electrodes which are made of conductive rubber After a headband was washed and dried up wire and electrodes must be reconnected as depicted on page 10 TECHNICAL SUPPORT Implies before the fact battery ...

Страница 11: ...heck battery connection replace battery or charge accumulator The Unit is being switched on auto checking procedure is in process TES session proceeds fine However during the session indicator Battery blinks Battery accumulator capacity started to run low during procedure Finish procedure and then replace battery or charge accumulator The Unit is being switched on auto checking procedure is in pro...

Страница 12: ... free of charge repair service or to replace Unit if repair is impossible Operating life 5 years since date of manufacture After warranty period has expired the Manufacturer will charge for providing with repair service Warranty and post warranty repair is performed by the Manufacturer or by certified district service centers ACCEPTANCE CERTIFICATE Unit DOCTOR TES 03 has been manufactured and cert...

Отзывы: