SPECIFICATIONS
Stimulating Current
................................................................
pulsed bipolar
Frequency, Hz........................................................................ 77.5
Current intensity, mA ............................................................ 0-1.5
Timer automatic, min................................................................. 30
Electric power, W ..................................................................... 6-9
Battery consumption current, mA .................................... up to 25
Weight, w/o battery, kg............................................................. 0.1
Dimensions, mm ......................................................60
х
114
х
52
STANDARD SET CONTAINS
:
unit, items................................................................................... 1
Headband, items......................................................................... 1
Frontal electrode, items............................................................... 1
Retromastoid electrodes, items................................................... 2
Power battery “Duracell” (6F22), items ....................................... 1
CD-disk with recorded psychotherapy session, items................. 1
User’s guide, items...................................................................... 1
Packing, items............................................................................. 1
22
COMMON TROUBLES AND REMEDIES
Revealed trouble
Probable cause
Remedies
The Unit is being
switched on, auto-
checking procedure is in
process, however LED
“Battery” is blinking
Battery / accumulator
capacity is running low
It may need to replace
battery or charge
accumulator
Unable to switch the Unit
on, autochecking
procedure is not started,
light indicators are off
Battery / accumulator is out
or not connected
Check battery connection,
replace battery or charge
accumulator
The Unit is being
switched on, auto-
checking procedure is in
process, TES session
proceeds fine. However,
during the session
indicator “Battery” blinks
Battery /accumulator
capacity started to run low
during procedure
Finish procedure, and then
replace battery or charge
accumulator
The Unit is being
switched on, auto-
checking procedure is in
process. However,
indicator “Battery” does
not highlight nor blink.
Intensity of current does
not increase
Wires and electrodes are not
connected or connected
improperly. Hairs or wires
could have been put
beneath electrodes
Check electrodes according
to the Paragraph 3. If
checking continues right,
put the electrodes back on.
Make sure no hairs or wires
are beneath them. In case
checking procedure fails –
disconnect them and
reconnect. Start checking
again. If failure repeats –
contact to Technical support
New battery / charged
accumulator is mounted,
but unable to switch the
Unit on, LEDs are not
highlighted. Unable to
perform a session
The Unit is out of order
Contact to the Technical
Support