Housing:
Powder-coated white CR steel cabinet with polyethylene leveling feet
Feed Air Pressure:
200 PSIG MAX
Feed Air Temperature: 40°F to 122°F
Nitrogen Outlet Flow: See graphs at right.
Prefilter:
One-micron absolute filtration. Dual glass micro-fiber filter beds coalesce and remove water
droplets. Remaining oil content: 1 ppm by weight. Includes slide indicator (to indicate need
for filter change), liquid level indicator and internal drain. ISO 8573.1 Quality Class - Solids:
Class 2, Oil Content: Class 4
Coalescing Filter:
Absolute filtration of 0.01µm particles; 99.999+% oil removal efficiency. Includes slide
indicator (to indicate need for filter change), liquid level indicator and internal drain. ISO
8573.1 Quality Class - Solids: Class 1, Oil Content: Class 1
Membrane Module:
Semipermeable hollow fiber bundles.
Electrical:
Power not needed to generate nitrogen, but low-pressure indicator light requires 120VAC.
User-supplied air compressor may require electricity.