TENSITRON STX-1 Series Скачать руководство пользователя страница 7

       STX-1 Series Instrument           

www.tensitron.com                      

pg. 7

 

 

Features of all Instruments 

 

Color graphic display:  

o

 

Large and easy to read, with adjustable backlighting. 

o

 

Shows Tension, Battery Charge Level, and Strap Material and 
Size Selected. 

o

 

Includes user-definable tension set points. 

 

Data Statistics: Monitor your process by viewing count, average, 
minimum/maximum, and standard deviation values of your logged data. 

 

Selectable Tension Units: Choose to display tension in Lb, Kg or daN. 

 

Up to ten calibrations can be programmed into the instrument. (Password-
protected to provide due diligence against unauthorized changes.) 

 

Adjustable LCD refresh rates allow for stable digital readings. 

 

Automatic shutoff after several minutes of non-use. 

 

All models are typically available from stock.  

 

All models are factory calibrated and ready for use.  

 

All calibration values are traceable to National Standards.  

 

Calibration certificate is included. 

 

Continuous operation while connected to power supply. 

 

Durable, lightweight carrying case with protective foam inserts included. 
(See Figure 2) 

 

 

 

 

 

Figure 2: Carrying case with instrument, calibration 

certificate, charger, adapters and operating instructions. 

Содержание STX-1 Series

Страница 1: ...STX 1 Series Instrument www tensitron com pg 1 Operating Instructions STX 1 Series Digital Strap Tension Meter...

Страница 2: ...2 Serial Output option 6 Features of all Instruments 7 5 BASIC OPERATION 8 Charging Instrument Batteries 8 Quick Start Instructions 8 Engaging Instrument onto Strap 10 6 DISPLAY SCREENS 10 SETUP Scree...

Страница 3: ...stions or need assistance please call us at 303 702 1980 or find us online at www tensitron com 1 WARRANTY POLICY STANDARD EQUIPMENT WARRANTY Tensitron warrants that all Tensitron manufactured equipme...

Страница 4: ...and PET strappings Power Requirements For Instrument Input 9 VDC 2 2A For AC Adapter 100 240V 50 60Hz 0 48A Power Supply operates with input voltages from 100 240V and includes several interchangeabl...

Страница 5: ...nstrument to do a job for which it is not recommended Check for damaged parts Inspect instrument before use Check for any binding of moving parts improper mountings broken parts and any other conditio...

Страница 6: ...nstruments include multiple selections for different strapping Models STXM 250 1 through STXM 1000 1 includes multiple selections for Poly and PET strappings The Model STXM 2000 1 includes selections...

Страница 7: ...calibrations can be programmed into the instrument Password protected to provide due diligence against unauthorized changes Adjustable LCD refresh rates allow for stable digital readings Automatic shu...

Страница 8: ...and corner of display Quick Start Instructions 1 Power unit on by pressing ON button Main display will indicate Tension Material Min and Max readings and other information See Figure 3 2 Select a scre...

Страница 9: ...n press ENTER b Next select from Kilograms DecaNewtons or LBS and then press ENTER 6 Select Material a Using the up or down arrows scroll to SELECT MATERIAL then press ENTER b Next scroll through the...

Страница 10: ...LCD refresh rates Scroll to SETUP and press ENTER Scroll to LCD REFRESH RATE and press ENTER Select your preference 1 2 or 5 Hz and press ENTER Note This feature is unrelated to Data Averaging o Data...

Страница 11: ...CHECK ACCURACY o Audio To turn the Beep ON or OFF Scroll to SETUP and press ENTER Scroll to AUDIO and press ENTER Select ON or OFF and press ENTER o Backlight To adjust the visual intensity of the LC...

Страница 12: ...ndard Deviation can be seen under the Data Logging View Data Stats screen To view this logged data Scroll to DATA LOGGING and press ENTER Scroll to VIEW DATA LOG and press ENTER o Clear Single Point L...

Страница 13: ...by holding it in the same attitude you ll be taking the readings and press the Enter or Zero button This will eliminate any small tare effects 3 Take readings along the strap FOR BEST RESULTS ALWAYS...

Страница 14: ...for it by following the instructions under Custom Names in Section 6 above 1 Scroll to SETUP and press ENTER Next scroll to RE CAL TENSION and press ENTER The screen will now display SETUP RE CALIBRAT...

Страница 15: ...m above and suspend known weights onto the strap in the values you ve previously selected when prompted See Figure 7 a CALIBRATE TENSION NUMBER OF POINTS __ ZERO UNIT Without any strap engaged to the...

Страница 16: ...733 S Bowen Street Longmont CO 80501 USA Phone 303 702 1980 Fax 303 702 1982 E mail sales tensitron com Web Site www tensitron com...