Proper Engagement of Instrument to Cable: 

 

Push the closure bar away from the instrument handle to raise 
the outer cable guides.  Then engage the instrument so that the 
tensioned cable is routed under the outer cable guides and 
over the center contact point.  Squeeze the closure bar shut 
until it locks into position beside the handle, and note the ten-
sion reading on the display.  In this example the tension is 41.0 
lb. 

Fig 3.

 

       

Fig 3, Proper Engagement

 

Page 14    ACX-Series Operating Instructions 

FOR BEST RESULTS ALWAYS TAKE SEVERAL READINGS OF THE 
CABLE TENSION BY DISENGAGING AND THEN REENGAGING THE 
INSTRUMENT FROM THE CABLE ITSELF. 
 

Note 1: Always disengage and then reengage the instrument to any cable 
whose tensions have been altered.  Failure to do so may cause readings to 
fall out of tolerance. 
 
Note 2:  You may find it helpful to store each reading by depressing the in-
strument’s “memory” button and then viewing the displayed average. 

Sec. 4.  ADDITIONAL TECHNICAL INFORMATION. 

 

 

1.   

SPEEDING UP OR SLOWING DOWN DISPLAY LCD REFRESH 

RATES - DAMPENING ADJUSTMENT.  

To either speed up or slow 

down the rate at which tension values refresh on the display, use the 

 

or 

 

keys, select 

SETUP

, and press 

ENTER.  

Next select 

Dampening 

and after entering this selection, choose the refresh rate 

from: 

1Hz, 2Hz or 5Hz

.  When new rate is highlighted, press 

ENTER

 
2. 

AUDIO.  

Turn on or off audio beep (with key inputs) by selecting:  

SETUP

, then select 

Audio

, and finally either select 

ON 

or 

OFF

 
 
 
 
 
 
 
 
 
 
 
 
3.   

BACKLIGHTING INTENSITY

.  Increase or decrease backlighting 

intensity by selecting:  

SETUP

, then 

BACKLIGHT

, and finally 

intensity level. 

 
4.   

DISPLAY CONTRAST

.  Increase or decrease LCD display contrast 

by selecting:  

SETUP

, then select 

CONTRAST

, and finally 

manipulate contrast using the 

 

or 

 

keys.  Remember to 

ENTER 

your new setting. 

 
5.   

VERSION

.  Instrument Model No. along with software version 

information can be viewed under 

SETUP

 and then selecting 

Version

 

6. 

CUSTOM CALIBRATIONS.  

Calibrations are password-protected to 

prevent unintended changes.  See Page 16 for step-by-step 
instructions, or contact your Metrology Department or Tensitron, Inc. 
for assistance.  

  

If you only want to verify your instrument’s accuracy refer to:  

How To Check Accuracy, on Page 12

ACX-Series Operating Instructions     Page 3 

Содержание ACX-Series

Страница 1: ...C 733 S Bowen Street Longmont CO 80501 USA Phone 303 702 1980 Fax 303 702 1982 E mail tensionmeters tensitron com Web Site www tensitron com Operating Instructions ACX Series Digital Aircraft Cable Te...

Страница 2: ...s for calibration take a sample length of your material and suspend it from above See Page 12 Fig 1 Next hang weights from your material in the weight values you previously selected in enter weights 1...

Страница 3: ...manual carefully to learn your tension meter s applications and limitations as well as the potential hazards associated with this type of instrument 2 AVOID DANGEROUS ENVIRONMENTS Do not use your inst...

Страница 4: ...oggle through the various selections using the or arrows until correct cable size is highlighted and press ENTER to make your selection If your specific material is not listed follow the calibration i...

Страница 5: ...REFRESH RATES DAMPENING ADJUSTMENT To either speed up or slow down the rate at which tension values refresh on the display use the or keys select SETUP and press ENTER Next select Dampening and after...

Страница 6: ...ext using the and arrows select ANALOG OUTPUT Press ENTER Using the and arrows select from ANALOG TENSION CURRENT or ANALOG TENSION VOLTAGE Once correct selection is made press ENTER 9 CALIBRATION OF...

Страница 7: ...units output serially will be the same units that are selected in the Tension Units menu The currently selected tension units such as lb kg or N are also displayed on the main tension screen e Perform...

Страница 8: ...ng stored data and can be adjusted to any setting b To retrieve or upload logged data from the instrument send a lower case d c Collected data will remain in memory regardless if instru ment is turned...

Страница 9: ...0 5 100 lb 0 5 lb 1 16 3 32 1 8 5 32 ACX 250 20 250 lb 0 5 lb 1 16 3 32 1 8 5 32 3 16 ACX 500 40 500 lb 1 0 lb 3 32 1 8 5 32 3 16 ACX CWT 20 250 lb 0 5 lb CWT 5 CWT 10 CWT 15 CWT 20 CWT 25 ACX CMB 20...

Страница 10: ...CWT 25 CWT Custom 9 Kilograms Decanewtons LBS Data Logging Optional Logging Duration Begin Logging View Data Setup Dampening Select 1 Hz 2 Hz 5 Hz Custom Names Resolution Select Custom 1 9 Re Calibrat...

Отзывы: