Step 2:

   Next you will be prompted 

ENTER WEIGHT 1.  

Using 

the 

 or 

 arrows select a minimum tension value. (

Do not use 

zero

.)  Press 

ENTER

.   

 
   Select increasing weight values for 

WEIGHT  2 THRU 4 

(No. 4 

being the maximum weight value).  At each weight number se-
lect your weight value using the 

 

and 

 

arrows.  

ENTER

 each 

selection.   

 

Next you will calibrate the instrument to the weight values you 

just selected.  To simulate tension loads for calibration, take a 
sample length of your material and suspend it from above.  
(See Page 12, Fig. 1)  Next hang weights from your material in 
the weight values you previously selected in “enter weights 1 
through 4”.   

 

  Your instrument will now prompt you

 PLACE WEIGHT 0.  

With 

zero tension applied to the instrument press 

ENTER.  

  

   Next you will be prompted 

PLACE  WEIGHT 1.  

Suspend the 

exact weight value from your material that you previously se-
lected.  Engage instrument to the tensioned material and then 
press 

ENTER.   

 

   Next suspend weight value 2 from your material, engage the 

instrument to your tensioned material and press 

ENTER.   

 

   Repeat procedure for steps 3 and 4.  Once you have entered 

the value for 

WEIGHT 4 

you have finished the calibration and 

will be returned to the Main Display.  The Main Display will now 
indicate the description of the calibration you just entered.  This 
signifies you have now selected the calibration you just per-
formed.   

 
   To change calibration use the 

 or 

 

arrows until you reach  

SELECT MATERIAL.  

Follow procedures outlined in Section 

3.6.   

 

Содержание ACX-Series

Страница 1: ...C 733 S Bowen Street Longmont CO 80501 USA Phone 303 702 1980 Fax 303 702 1982 E mail tensionmeters tensitron com Web Site www tensitron com Operating Instructions ACX Series Digital Aircraft Cable Te...

Страница 2: ...s for calibration take a sample length of your material and suspend it from above See Page 12 Fig 1 Next hang weights from your material in the weight values you previously selected in enter weights 1...

Страница 3: ...manual carefully to learn your tension meter s applications and limitations as well as the potential hazards associated with this type of instrument 2 AVOID DANGEROUS ENVIRONMENTS Do not use your inst...

Страница 4: ...oggle through the various selections using the or arrows until correct cable size is highlighted and press ENTER to make your selection If your specific material is not listed follow the calibration i...

Страница 5: ...REFRESH RATES DAMPENING ADJUSTMENT To either speed up or slow down the rate at which tension values refresh on the display use the or keys select SETUP and press ENTER Next select Dampening and after...

Страница 6: ...ext using the and arrows select ANALOG OUTPUT Press ENTER Using the and arrows select from ANALOG TENSION CURRENT or ANALOG TENSION VOLTAGE Once correct selection is made press ENTER 9 CALIBRATION OF...

Страница 7: ...units output serially will be the same units that are selected in the Tension Units menu The currently selected tension units such as lb kg or N are also displayed on the main tension screen e Perform...

Страница 8: ...ng stored data and can be adjusted to any setting b To retrieve or upload logged data from the instrument send a lower case d c Collected data will remain in memory regardless if instru ment is turned...

Страница 9: ...0 5 100 lb 0 5 lb 1 16 3 32 1 8 5 32 ACX 250 20 250 lb 0 5 lb 1 16 3 32 1 8 5 32 3 16 ACX 500 40 500 lb 1 0 lb 3 32 1 8 5 32 3 16 ACX CWT 20 250 lb 0 5 lb CWT 5 CWT 10 CWT 15 CWT 20 CWT 25 ACX CMB 20...

Страница 10: ...CWT 25 CWT Custom 9 Kilograms Decanewtons LBS Data Logging Optional Logging Duration Begin Logging View Data Setup Dampening Select 1 Hz 2 Hz 5 Hz Custom Names Resolution Select Custom 1 9 Re Calibrat...

Отзывы: