12. GENERAL PAD ADVICE (cont)

 

The pads provided are latex –free.

 

Replace pads when they lose their stickiness.  

 

Poor connection may cause discomfort and skin irritation

 

Storage life of an unopened pack of electrodes is 2 years.  

 

This may be affected by very high temperatures or very  

 

low humidity.

13. PAD ALARM

Perfect pad contact and condition is important to obtain maximum 

benefit from any TENS unit. To ensure best performance the 

“itouch easy”

 has been fitted with a “Pad Alarm”. 

If the contact between any of the pads and the skin or the quality 

of the pads is below the required standard the pad alarm will be 

triggered.  An audible bleep will be heard and at the top of the 

screen the word “PADS” will flash.  

 

When the pad alarm is triggered, the Comfort Strength Control 

facility will automatically lower the intensity to a comfortable level 

thereby avoiding the possibility of any electrical shock, however 

small. If the alarm is triggered due to the pads coming partly away 

from the body, then when they are put back the intensity will 

automatically return to the original level.

The “PADS” alarm does not function on either channel when the 

intensity levels are below 5.0mA.

Содержание itouch easy

Страница 1: ...c t i o n s f o r U s e I N T E L L I G E N T T E N S TensCare Ltd 9 Blenheim Road Longmead Business Park Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 Email painaway tenscare co uk www tenscare co uk...

Страница 2: ...Screen Information 9 8 Contents 15 9 Installation of Batteries 15 10 Connecting Mono Lead wire 17 11 Positioning of Electrode Pads 18 12 General Pad Advice 20 13 Pad Alarm 22 14 The Belt Cradle 23 15...

Страница 3: ...th CSC you can increase the strength on the itouch easy in very small steps which is much more comfortable than conventional TENS units One Touch Memory Start The itouch easy has a memory which allows...

Страница 4: ...ificant reduction in pain if not complete pain relief 4 SIDE EFFECTS There are no known side effects to TENS use and long term TENS use is not harmful 5 THE PROGRAMMES There is no one programme approp...

Страница 5: ...sation Specific Uses E Constant This is similar to A but has Most types of acute deeper penetration and a and chronic pain stronger feel It is more including arthritis likely to produce muscle sciatic...

Страница 6: ...his enables the body to receive many different signals and can be very beneficial and lessen any effect of accommodation See programmes F and G Han Stimulation Professor Han in China found that altern...

Страница 7: ...ton Channel 2 Strength UP Channel 2 Strength DOWN Channel 1 Strength UP Time Selector Channel 2 Strength UP 7 SCREEN INFORMATION Programme selected Frequency selected Channel 1 strength selected Pulse...

Страница 8: ...to switch the unit off press and hold down this button for 3 seconds ON Channel 1 Strength UP button Note Channel 1 controls the output through the white leads To switch unit on Press and hold down fo...

Страница 9: ...press the Programme Selector button and release The programmes are shown as A B C D E F G or H Time Selector button Each time the unit is switched on it defaults back to continuous mode known as TIMER...

Страница 10: ...l Ensure that the batteries are inserted the right way as shown in battery compartment When the batteries are running low a low battery indicator will show on the screen battery symbol and it is impor...

Страница 11: ...the other end of the Mono Lead wire the lead wire splits into four pin ends 2 white and 2 black Push the pin ends into the pigtail ends of the electrode pads The mono lead wire can be split as requir...

Страница 12: ...rotective plastic shield by holding and lifting one corner of the pad and pulling Do not pull on the pigtail wire of the pad After use always replace the pads on the plastic liner and replace in the r...

Страница 13: ...contact between any of the pads and the skin or the quality of the pads is below the required standard the pad alarm will be triggered An audible bleep will be heard and at the top of the screen the...

Страница 14: ...ng solution than soap and water Clean the case and lead wires at least once a week by wiping with a damp cloth and a solution of mild soap and water Wipe dry 16 TROUBLESHOOTING Problem Possible causes...

Страница 15: ...S may affect seizure threshold During the first three months of pregnancy It is not known whether TENS may affect foetal development On the abdomen in the later stages of pregnancy Stop use immediatel...

Страница 16: ...brain are unknown Near malignant tumours In vitro experiments have shown that electricity can promote cell growth Do not Ignore any allergic reaction to the electrode pads If a skin irritation develo...

Страница 17: ...h 19 CONSUMABLES AND SERVICING Replacement electrode pads new batteries and lead wires are available from your supplier or distributor see back cover for contact details by mail order from TensCare by...

Страница 18: ...ked with a crossed out wheelie bin symbol as depicted below To comply with the Directive you can return your old electro therapy unit to us for disposal Simply print a postage paid PACKETPOST RETURNS...

Страница 19: ...ould be protected from liquids NB The electrical specifications are nominal and subject to variation from the listed values due to normal production tolerances EC Declaration of Conformity TensCare Lt...

Страница 20: ...5 UT 95 dip in Not applicable Not applicable interruptions and UT for 0 5 cycle voltage variations 40 UT 60 dip in on power supply UT for 5 cycles input lines 70 UT 30 dip in IEC 61000 4 11 UT for 25...

Страница 21: ...marked with the following symbol Rated maximum output power of transmitter W NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations El...

Страница 22: ...39 NOTES...

Отзывы: